Proper theoretical analysis of the oxygen wash-in kinetics of circle breathing systems
- PMID: 22252241
- DOI: 10.1111/j.1467-2995.2011.00674.x
Proper theoretical analysis of the oxygen wash-in kinetics of circle breathing systems
Abstract
Objective: To correctly define the theoretical treatment of oxygen wash-in kinetics in circle breathing systems and reevaluate previously published results for the rate of change of oxygen concentration in a large animal circle breathing system.
Study design: Theoretical analysis of previously published data.
Animals: None.
Methods: Previously published data for the rate of change of oxygen concentration in a large animal circle breathing system with two reservoir bag sizes (40 and 20 L) at three different flow rates (3, 6, and 10 L minute(-1)) were obtained from the original publication, via digital extraction, and analyzed to determine system time constants. The results of this analysis were compared to those originally reported and the level of mathematical agreement between experiment and theory was quantified.
Results: Theoretical time constants for the system with 40 L reservoir bag with flow rates of 3, 6, and 10 L minute(-1) were 16.4, 8.2, and 4.9 minutes, respectively. Experimentally derived time constants for this system with flow rates of 3, 6, and 10 L minute(-1) were 18.1, 9.2, and 5.4 minutes, respectively. Percent differences between experimental and theoretical time constants for this system with flow rates of 3, 6, and 10 L minute(-1) were 10.4, 12.2, and 10.2%, respectively. For the system with a 20 L reservoir bag and 6 L minute(-1) flow rate, the theoretical and experimentally derived time constants were 5.5 and 5.6 minutes, respectively, with a 1.8% difference. The average relative deviations between theory and experiment for the system at 6 L minute(-1) flow with a 40 or 20 L reservoir bag were 1.3% and 3.0%, respectively.
Conclusions and clinical relevance: Proper theoretical analysis of experimentally obtained data for the wash-in kinetics of oxygen into a large animal circle breathing system leads to improved mathematical agreement between theory and experiment when compared to the originally published results. Application of this method should allow more accurate prediction of the rate of change of oxygen concentration in anesthetic circuits.
© 2012 The Authors. Veterinary Anaesthesia and Analgesia. © 2012 Association of Veterinary Anaesthetists and the American College of Veterinary Anesthesiologists.
Similar articles
-
Rate of change of oxygen concentration for a large animal circle anesthetic system.Am J Vet Res. 2005 Oct;66(10):1675-8. doi: 10.2460/ajvr.2005.66.1675. Am J Vet Res. 2005. PMID: 16273896
-
Preoxygenation with the Mapleson D system requires higher oxygen flows than Mapleson A or circle systems.Can J Anaesth. 2007 Feb;54(2):141-5. doi: 10.1007/BF03022011. Can J Anaesth. 2007. PMID: 17272254
-
[Anesthetic gas consumption and costs in a closed system with the PhysioFlex anesthesia equipment].Anaesthesist. 1993 Sep;42(9):638-43. Anaesthesist. 1993. PMID: 8214536 German.
-
[Anesthesia with low fresh gas flow in clinical routine use].Anasthesiol Intensivmed Notfallmed Schmerzther. 1995 Aug;30(5):268-75. doi: 10.1055/s-2007-996491. Anasthesiol Intensivmed Notfallmed Schmerzther. 1995. PMID: 7548477 Review. German.
-
The case for rebreathing circuits for very small animals.Vet Clin North Am Small Anim Pract. 1992 Mar;22(2):400-3. doi: 10.1016/s0195-5616(92)50650-7. Vet Clin North Am Small Anim Pract. 1992. PMID: 1585586 Review.
Cited by
-
The Modification and Performance of a Large Animal Anesthesia Machine (Tafonius®) in Order to Deliver Xenon to a Horse.Front Vet Sci. 2017 Sep 29;4:162. doi: 10.3389/fvets.2017.00162. eCollection 2017. Front Vet Sci. 2017. PMID: 29034250 Free PMC article.
-
Do the Manual or Computer-Controlled Flowmeters Generate Similar Isoflurane Concentrations in Tafonius?Front Vet Sci. 2019 May 29;6:160. doi: 10.3389/fvets.2019.00160. eCollection 2019. Front Vet Sci. 2019. PMID: 31192237 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources