. 2012 Apr;7(1):45-50.
doi: 10.1007/s11751-012-0127-6. Epub 2012 Jan 26.

Union of an intra-articular distal radius fracture after successive failures of three locking plates: a case report

Affiliations

Union of an intra-articular distal radius fracture after successive failures of three locking plates: a case report

S K Khan et al. Strategies Trauma Limb Reconstr. 2012 Apr.

Abstract

We report a case of a 30-year old male, who presented with a right distal radius intra-articular fracture complicated by compartment syndrome. He was treated with fasciotomies and fracture fixation with a 3.5 mm LCP (Synthes(™)), followed 7 days later by skin graft. Repeat radiographs 8 weeks later showed a break across the plate at the level of an unfilled screw hole over the fracture. He underwent exchange plating with a 2.4 mm LCP Distal Radius Plate (Synthes(™)). This revision was complicated by an infected wound dehiscence 2 weeks later requiring multiple procedures. Radiographs at 20 weeks showed broken distal screws. A second revision was performed. At 12 months, the fracture had healed clinically and radiologically, but the three distal screws had broken. We discuss the multifactorial failures of the these three attempts at osteosynthesis, and which factors helped achieve osseous union. We also discuss the literature on volar locking plate breakage and conclude with the recommendations to avoid this rare complication.

PubMed Disclaimer

Figures

Fig. 1
Fig. 1
Radiographs after injury, showing the distal radius fracture
Fig. 2
Fig. 2
Radiographs at 6 weeks after the index procedure. Broken plate
Fig. 3
Fig. 3
Radiographs at 20 weeks after the first revision procedure
Fig. 4
Fig. 4
Radiographs at 26 weeks after the first revision procedure. Broken screws
Fig. 5
Fig. 5
Radiographs at 6 weeks after the second revision procedure (with bone graft)
Fig. 6
Fig. 6
Clinical examination at 6 weeks after the second revision
Fig. 7
Fig. 7
Radiographs at 1 year after second revision
Fig. 8
Fig. 8
Suggested causes for sequential failures of the three plates

References

    1. Downing ND, Karantana A. A revolution in the management of fractures of the distal radius? J Bone Joint Surg Br. 2008;90(10):1271–1275. doi: 10.1302/0301-620X.90B10.21293. - DOI - PubMed
    1. Al-Rashid M, Theivendran K, Craigen MA. Delayed ruptures of the extensor tendon secondary to the use of volar locking compression plates for distal radial fractures. J Bone Joint Surg Br. 2006;88(12):1610–1612. doi: 10.1302/0301-620X.88B12.17696. - DOI - PubMed
    1. Arora R, Lutz M, Hennerbichler A, Krappinger D, Espen D, Gabl M. Complications following internal fixation of unstable distal radius fracture with a palmar locking-plate. J Orthop Trauma. 2007;21(5):316–322. doi: 10.1097/BOT.0b013e318059b993. - DOI - PubMed
    1. Rozental TD, Blazar PE. Functional outcome and complications after volar plating for dorsally displaced, unstable fractures of the distal radius. J Hand Surg Am. 2006;31(3):359–365. doi: 10.1016/j.jhsa.2005.10.010. - DOI - PubMed
    1. Berglund LM, Messer TM. Complications of volar plate fixation for managing distal radius fractures. J Am Acad Orthop Surg. 2009;17(6):369–377. - PubMed

LinkOut - more resources