Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2012 Mar 25:62:135-9.
doi: 10.1016/j.jpba.2012.01.003. Epub 2012 Jan 14.

An improved simple LC-MS/MS method for the measurement of serum aripiprazole and its major metabolite

Affiliations

An improved simple LC-MS/MS method for the measurement of serum aripiprazole and its major metabolite

Matteo Caloro et al. J Pharm Biomed Anal. .

Abstract

Background and objectives: Current liquid chromatographic tandem mass spectrometry (LC-MS/MS) methods to measure serum levels of aripiprazole (Ar) and dehydroaripiprazole (DHAr) are sensitive, but difficult to use in a hospital context. We aimed to develop a rapid LC-MS/MS method allowing reliable level measurement in the presence of co-administered drugs, withdrawing samples from 22 patients with acute agitation receiving 9.75 mg aripiprazole IM injection.

Method: We developed a sensitive and selective HPLC-MS/MS method to measure serum Ar and DHAr levels in a hospital laboratory, requiring minimal sample preparation and inferior sample volume compared to previous LC-MS/MS methods. Analytes were separated on a reversed-phase HPLC (run-time, 10 min). A triple quadrupole tandem mass spectrometer was used for quantitative analysis in positive mode by a multiple reaction monitoring. Samples were drawn 2, 4, 6, and 24h post-injection.

Results: Calibration curves (2-1000 ng/mL for Ar and 3.5-500 ng/mL for DHAr) were linear, with mean correlation coefficient >0.9998. Within- and between-day precision and accuracy were within 10%. Mean recovery was 95.2 ± 4.5% for Ar and 97.6 ± 7.2% for DHAr. Ar and DHAr peaks were not affected by other co-administered psychotropic drugs.

Conclusion: Our method measured Ar and DHAr concentrations reliably, simply and rapidly without employing many reagents, as currently existing methods.