Review
. 2012 Feb;20(2):63-8.
doi: 10.5435/JAAOS-20-02-063.

Nerve conduits for nerve repair or reconstruction

Affiliations
Review

Nerve conduits for nerve repair or reconstruction

D Nicole Deal et al. J Am Acad Orthop Surg. 2012 Feb.

Abstract

Advances in treating peripheral nerve lesions have resulted from research in nerve regeneration and the use biomaterials as well as synthetic materials. When direct tensionless repair of peripheral nerve lesions is not possible, nerve conduits may be used to bridge digital sensory nerve gaps of ≤3 cm. Nerve autograft is the benchmark for larger, longer, mixed, or motor nerve defects. Biologic, autogenous conduits-typically veins or, rarely, arteries-have demonstrated their utility in nerve gaps <3 cm in length. Three types of bioabsorbable conduit have been approved by the US Food and Drug Administration, constructed of collagen, polyglycolic acid, or caprolactone. Caprolactone conduits have been found to be equivalent in results to autograft. Collagen conduits are next best, and polyglycolic acid conduits are functionally inferior. Further research and prospective, multicenter, large-scale trials are needed to help establish the role of synthetic, bioabsorbable conduits in peripheral nerve reconstruction.

PubMed Disclaimer