Validation of a new automatic system for biplane quantitative coronary arteriography
- PMID: 2230301
- DOI: 10.1007/BF01833978
Validation of a new automatic system for biplane quantitative coronary arteriography
Abstract
In a collaboration between the University of Texas (software) and the University of Zürich (hardware) a compact, automatic system for biplane quantitative coronary arteriography was developed. The system is based on a 35 mm film projector, a slow-scan CCD-camera (image digitizing) and a computer workstation (Apollo DN 3000, image storage and processing). A new calibration procedure based on two fixed reference points in the center of the image intensifier was used (isocenter technique). Contour detection of coronary arteries was carried out in biplane projection using a geometric-densitometric edge-detection algorithm. The proximal and distal luminal areas, as well as the minimal luminal area of the stenotic vessel segment were determined. Accuracy and precision were determined from precision drilled holes in a plexiglas cube which were filled with 50%, 75% and 100% contrast medium. The diameter of the holes ranged from 0.5 to 5.0 mm. The mean difference and the standard deviation of the differences between the true and the measured diameters were 0.12 +/- 0.14 mm for plane A and 0.26 +/- 0.17 mm for plane B, respectively. After a second order correction the mean difference amounted to 0.02 +/- 0.09 mm for plane A and 0.02 +/- 0.12 mm for plane B, respectively. Intra- and interobserver variability were evaluated in 5 patients (age 60 +/- 10 years) with coronary artery disease using 16 normal and 5 stenotic vessel segments (cross-sectional area ranging from 0.8 to 8.7 mm2). Two independent observers analyzed the same vessel segment twice. Intraobserver variability expressed as the standard error of estimate in percent of the mean angiographic vessel area (SEE) amounted to 2.1% for observer 1 and 4.4% for observer 2, respectively. Interobserver variability expressed as SEE was 4.1% for measurement 1 and 3.6% for measurement 2, respectively.
Similar articles
-
Automated quantitative coronary arteriography: morphologic and physiologic validation in vivo of a rapid digital angiographic method.Circulation. 1987 Feb;75(2):452-60. doi: 10.1161/01.cir.75.2.452. Circulation. 1987. PMID: 3802448
-
A new method for automatic identification of coronary arteries in standard biplane angiograms.Int J Card Imaging. 1994 Dec;10(4):253-61. doi: 10.1007/BF01137716. Int J Card Imaging. 1994. PMID: 7722346
-
Accuracy and precision of quantitative digital coronary arteriography: observer-, short-, and medium-term variabilities.Cathet Cardiovasc Diagn. 1993 Mar;28(3):187-98. doi: 10.1002/ccd.1810280301. Cathet Cardiovasc Diagn. 1993. PMID: 8439993
-
Assessment of the vasomotility of epicardial coronary arteries with quantitative coronary angiography.Z Kardiol. 1989;78 Suppl 6:143-8. Z Kardiol. 1989. PMID: 2694662 Review.
-
Definition and measurement of restenosis after successful coronary angioplasty: implications for clinical trials.Am Heart J. 1993 May;125(5 Pt 1):1394-408. doi: 10.1016/0002-8703(93)91013-5. Am Heart J. 1993. PMID: 8480594 Review.
Cited by
-
Towards a geometrically correct 3-D reconstruction of tortuous coronary arteries based on biplane angiography and intravascular ultrasound.Int J Card Imaging. 1997 Dec;13(6):451-62. doi: 10.1023/a:1005843222820. Int J Card Imaging. 1997. PMID: 9415847
-
The impact of vessel and catheter position on the measurement accuracy in catheter-based quantitative coronary angiography.Int J Card Imaging. 1998 Aug;14(4):217-27. doi: 10.1023/a:1006067117225. Int J Card Imaging. 1998. PMID: 9934610
-
A system for determination of 3D vessel tree centerlines from biplane images.Int J Card Imaging. 2000 Oct;16(5):315-30. doi: 10.1023/a:1026528209003. Int J Card Imaging. 2000. PMID: 11215917
-
Biplane X-ray angiograms, intravascular ultrasound, and 3D visualization of coronary vessels.Int J Card Imaging. 1999 Dec;15(6):495-512. doi: 10.1023/a:1006372704091. Int J Card Imaging. 1999. PMID: 10768744 Review.