Thin CaSO4:Dy thermoluminescent dosimeters for calibration of 90Sr+90Y applicators
- PMID: 22304964
- DOI: 10.1016/j.apradiso.2012.01.003
Thin CaSO4:Dy thermoluminescent dosimeters for calibration of 90Sr+90Y applicators
Abstract
Clinical applicators are used in brachytherapy to treat superficial lesions of skin and eye. They should be periodically calibrated according to quality control programs and international recommendations. Thin CaSO(4):Dy thermoluminescent dosimeters were used to calibrate various applicators with a dermatological applicator as a reference. The obtained absorbed dose rates were compared with those quoted in their calibration certificates. Depth-dose curves were constructed for all the applicators. A mail dosimetry system was developed for calibration of clinical applicators.
Copyright © 2012 Elsevier Ltd. All rights reserved.