. 1990 Oct;6(4):276-83.
doi: 10.1007/BF02842487.

Rib cage and abdominal piezoelectric film belts to measure ventilatory airflow

Affiliations

Rib cage and abdominal piezoelectric film belts to measure ventilatory airflow

B E Pennock. J Clin Monit. 1990 Oct.

Abstract

Piezoelectric film-based respiratory belts are described and tested. To the extent that a two-degree of freedom model consisting of rib cage and abdominal motion is able to assist in quantifying ventilation, the piezoelectric belts can monitor flow in a manner analogous to the monitoring of volume with two magnetometer pairs or two respiratory inductive plethysmograph belts. The piezoelectric belts are shown to measure flow linearly when compared with a screen pneumotachometer to a flow of at least 2.6 L/s. There is no phase shift between the peak flow of belts and the pneumotachometer up to a frequency of at least 9.2 Hz. During normal ventilation, 68 and 95% of the peak flows measured with the belts fall within +/- 10 and 20%, respectively, of the flows measured with a screen pneumotachometer.

PubMed Disclaimer

Similar articles

Cited by

References

    1. J Appl Physiol (1985). 1986 Mar;60(3):928-34 - PubMed
    1. J Appl Physiol. 1967 Mar;22(3):407-22 - PubMed
    1. J Appl Physiol Respir Environ Exerc Physiol. 1979 Feb;46(2):399-406 - PubMed
    1. Bull Eur Physiopathol Respir. 1982 Jul-Aug;18(4):643-58 - PubMed
    1. Science. 1967 Jun 9;156(3780):1383-4 - PubMed

MeSH terms