Accuracy of four indirect methods of blood pressure measurement, with hemodynamic correlations
- PMID: 2230858
- DOI: 10.1007/BF02842488
Accuracy of four indirect methods of blood pressure measurement, with hemodynamic correlations
Abstract
In 38 adults undergoing cardiac surgery, 4 indirect blood pressure techniques were compared with brachial arterial blood pressure at predetermined intervals before and after cardiopulmonary bypass. Indirect blood pressure measurement techniques included automated oscillometry, manual auscultation, visual onset of oscillation (flicker) and return-to-flow methods. Hemodynamic measurements or calculations included heart rate, cardiac index, stroke volume index, and systemic vascular resistance index. Indirect and intraarterial blood pressure values were compared by simple linear regression by patient and measurement period. Measurement errors (arterial minus indirect blood pressure) were calculated, and stepwise regression assessed the relationship between measurement error and heart rate, cardiac index, stroke volume index, and systemic vascular resistance index. Indirect to intraarterial blood pressure correlation coefficients varied over time, with the strongest correlations often occurring at the first and last measurement periods (preinduction and 60 minutes after cardiopulmonary bypass), particularly for systolic blood pressure. Within-patient correlations between indirect and arterial blood pressure varied widely--they were consistently high or low in some patients. In other patients, correlations were especially weak with a particular indirect blood pressure method for systolic, mean, or diastolic blood pressure; in some cases indirect blood pressure was inadequate for clinical diagnosis of acute blood pressure changes or trends. The mean correlations between indirect and direct blood pressure values were, for systolic blood pressure: 0.69 for oscillometry, 0.77 for auscultation, 0.73 for flicker, and 0.74 for return-to-flow; for mean blood pressure: 0.70 for oscillometry and 0.73 for auscultation; and for diastolic blood pressure: 0.73 for oscillometry and 0.69 for auscultation. The mean measurement errors (arterial minus indirect values) for the individual indirect blood pressure methods were, for systolic: 0 mm Hg for oscillometry, 9 mm Hg for auscultation, -5 mm Hg for flicker, 7 mm Hg for return-to-flow; for mean: -6 mm Hg for oscillometry, and -3 mm Hg for auscultation; and for diastolic: -9 mm Hg for oscillometry and -8 mm Hg for auscultation. Mean measurement error for systolic blood pressure was thus least with automated oscillometry and greatest with manual auscultation, while standard deviations ranging from 9 to 15 mm Hg confirmed the highly variable nature of single indirect blood pressure measurements. Except for oscillometric diastolic blood pressure, a combination of systemic hemodynamics (heart rate, stroke volume index, systemic vascular resistance index, and cardiac index) correlated with each indirect blood pressure measurement error, which suggests that particular numeric ranges of these variables minimize measurement error.(ABSTRACT TRUNCATED AT 400 WORDS)
Similar articles
-
Radial artery tonometry: moderately accurate but unpredictable technique of continuous non-invasive arterial pressure measurement.Br J Anaesth. 1996 Mar;76(3):405-11. doi: 10.1093/bja/76.3.405. Br J Anaesth. 1996. PMID: 8785142 Clinical Trial.
-
Noninvasive estimation of central aortic pressure using the oscillometric method for analyzing systemic artery pulsatile blood flow: comparative study of indirect systolic, diastolic, and mean brachial artery pressure with simultaneous direct ascending aortic pressure measurements.Am Heart J. 1982 May;103(5):879-86. doi: 10.1016/0002-8703(82)90403-3. Am Heart J. 1982. PMID: 7072592
-
Assessment of the accuracy of indirect blood pressure measurements.Jpn Heart J. 1997 May;38(3):393-407. doi: 10.1536/ihj.38.393. Jpn Heart J. 1997. PMID: 9290574
-
From Korotkoff and Marey to automatic non-invasive oscillometric blood pressure measurement: does easiness come with reliability?Expert Rev Med Devices. 2016;13(2):179-89. doi: 10.1586/17434440.2016.1128821. Epub 2016 Jan 8. Expert Rev Med Devices. 2016. PMID: 26641026 Review.
-
Peripheral vascular effects on auscultatory blood pressure measurement.J Clin Monit. 1993 Jan;9(1):9-17. doi: 10.1007/BF01627631. J Clin Monit. 1993. PMID: 8463815 Review.
Cited by
-
Theoretical analysis of non-invasive oscillometric maximum amplitude algorithm for estimating mean blood pressure.Med Biol Eng Comput. 1997 May;35(3):271-8. doi: 10.1007/BF02530049. Med Biol Eng Comput. 1997. PMID: 9246863
-
Effect of hydroxyethyl starch on postoperative kidney function in patients having noncardiac surgery.Anesthesiology. 2014 Oct;121(4):730-9. doi: 10.1097/ALN.0000000000000375. Anesthesiology. 2014. PMID: 25054470 Free PMC article.
-
Continuous finger arterial pressure: utility in the cardiovascular laboratory.Clin Auton Res. 1991 Mar;1(1):43-53. doi: 10.1007/BF01826057. Clin Auton Res. 1991. PMID: 1821665
-
Monitoring during paediatric cardiac anaesthesia.Can J Anaesth. 1994 Sep;41(9):818-44. doi: 10.1007/BF03011590. Can J Anaesth. 1994. PMID: 7955000 Review.
-
Evaluation of non-invasive blood pressure measurement by the Finapres method at rest and during dynamic exercise in subjects with cardiovascular insufficiency.Clin Auton Res. 1994 Apr;4(1-2):49-56. doi: 10.1007/BF01828838. Clin Auton Res. 1994. PMID: 8054837
References
MeSH terms
LinkOut - more resources
Other Literature Sources
Medical