Determination of absorbed dose to water around a clinical HDR (192)Ir source using LiF:Mg,Ti TLDs demonstrates an LET dependence of detector response
- PMID: 22320824
- DOI: 10.1118/1.3675401
Determination of absorbed dose to water around a clinical HDR (192)Ir source using LiF:Mg,Ti TLDs demonstrates an LET dependence of detector response
Abstract
Purpose: Experimental radiation dosimetry with thermoluminescent dosimeters (TLDs), calibrated in a (60)Co or megavoltage (MV) photon beam, is recommended by AAPM TG-43U1for verification of Monte Carlo calculated absorbed doses around brachytherapy sources. However, it has been shown by Carlsson Tedgren et al. [Med. Phys. 38, 5539-5550 (2011)] that for TLDs of LiF:Mg,Ti, detector response was 4% higher in a (137)Cs beam than in a (60)Co one. The aim of this work was to investigate if similar over-response exists when measuring absorbed dose to water around (192)Ir sources, using LiF:Mg,Ti dosimeters calibrated in a 6 MV photon beam.
Methods: LiF dosimeters were calibrated to measure absorbed dose to water in a 6 MV photon beam and used to measure absorbed dose to water at distances of 3, 5, and 7 cm from a clinical high dose rate (HDR) (192)Ir source in a polymethylmethacrylate (PMMA) phantom. Measured values were compared to values of absorbed dose to water calculated using a treatment planning system (TPS) including corrections for the difference in energy absorption properties between calibration quality and the quality in the users' (192)Ir beam and for the use of a PMMA phantom instead of the water phantom underlying dose calculations in the TPS.
Results: Measured absorbed doses to water around the (192)Ir source were overestimated by 5% compared to those calculated by the TPS. Corresponding absorbed doses to water measured in a previous work with lithium formate electron paramagnetic resonance (EPR) dosimeters by Antonovic et al. [Med. Phys. 36, 2236-2247 (2009)], using the same irradiation setup and calibration procedure as in this work, were 2% lower than those calculated by the TPS. The results obtained in the measurements in this work and those obtained using the EPR lithium formate dosimeters were, within the expanded (k = 2) uncertainty, in agreement with the values derived by the TPS. The discrepancy between the results using LiF:Mg,Ti TLDs and the EPR lithium formate dosimeters was, however, statistically significant and in agreement with the difference in relative detector responses found for the two detector systems by Carlsson Tedgren et al. [Med. Phys. 38, 5539-5550 (2011)] and by Adolfsson et al. [Med. Phys. 37, 4946-4959 (2010)].
Conclusions: When calibrated in (60)Co or MV photon beams, correction for the linear energy transfer (LET) dependence of LiF:Mg,Ti detector response will be needed as to measure absorbed doses to water in a (192)Ir beam with highest accuracy. Such corrections will depend on the manufacturing process (MTS-N Poland or Harshaw TLD-100) and details of the annealing and read-out schemes used.
Similar articles
-
Response of LiF:Mg,Ti thermoluminescent dosimeters at photon energies relevant to the dosimetry of brachytherapy (<1 MeV).Med Phys. 2011 Oct;38(10):5539-50. doi: 10.1118/1.3633892. Med Phys. 2011. PMID: 21992372
-
Using LiF:Mg,Cu,P TLDs to estimate the absorbed dose to water in liquid water around an 192Ir brachytherapy source.Med Phys. 2014 Jan;41(1):011711. doi: 10.1118/1.4851636. Med Phys. 2014. PMID: 24387503
-
Suitability of microDiamond detectors for the determination of absorbed dose to water around high-dose-rate 192 Ir brachytherapy sources.Med Phys. 2018 Jan;45(1):429-437. doi: 10.1002/mp.12694. Epub 2017 Dec 16. Med Phys. 2018. PMID: 29171060
-
Determination of the dose rate around a HDR 192Ir brachytherapy source with the microDiamond and the microSilicon detector.Z Med Phys. 2023 Nov;33(4):463-478. doi: 10.1016/j.zemedi.2022.07.004. Epub 2022 Aug 26. Z Med Phys. 2023. PMID: 36038432 Free PMC article. Review.
-
Response of thermoluminescent detectors to charged particles and to neutrons.Radiat Meas. 2004 Aug-Dec;38(4-6):407-12. doi: 10.1016/j.radmeas.2003.12.023. Radiat Meas. 2004. PMID: 15856571 Review.
Cited by
-
Experimental Determination of Radial Dose Function and Anisotropy Function of GammaMed Plus 192Ir High-Dose-Rate Brachytherapy Source in a Bounded Water Phantom and its Comparison with egs_brachy Monte Carlo Simulation.J Med Phys. 2019 Oct-Dec;44(4):246-253. doi: 10.4103/jmp.JMP_60_19. Epub 2019 Dec 11. J Med Phys. 2019. PMID: 31908383 Free PMC article.
-
Comparison of TLD calibration methods for 192Ir dosimetry.J Appl Clin Med Phys. 2013 Jan 7;14(1):4037. doi: 10.1120/jacmp.v14i1.4037. J Appl Clin Med Phys. 2013. PMID: 23318392 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Miscellaneous