Resonance frequency analysis-reliability in third generation instruments: Osstell mentor®
- PMID: 22322519
- PMCID: PMC3482525
- DOI: 10.4317/medoral.17861
Resonance frequency analysis-reliability in third generation instruments: Osstell mentor®
Abstract
Few studies assess repeatability and reproducibility in registers of resonance frequency analysis (a value of dental implant stability).
Objective: Few studies assess repeatability and reproducibility in resonance frequency analyses (implant stability evaluation). This study is aimed at assessing reliability (repeatabilty and reproducibility) in the Osstell Mentor® system using the intraclass correlation coefficient (ICC) as the statistical method.
Study design: ISQ measurements of RFA were carried out by means of the Osstell Mentor® instrument in 58 implants in 19 patients. Six measurements were performed on each implant by means of two different Smart-Pegs (I and II). Three consecutive measurements were registered with each transducer.
Results: Average ISQ varied from 72.43 to 72.60 and 73.26 in the first, second and third measurements, respectively with the SamrtPeg I and from 72.98 to 73.26 and 73.74 in the first, second and third measurements, respectively with the SamrtPeg II . Exactly equal values were observed in 10.43 and 12.1% of the cases with Smart-Pegs I and II, respectively. The intraclass correlation coefficient was 0.96 and 0.96 for Smart Pegs I and II, respectively. Repeatability and reproducibility was 0.97 for both Smart-Pegs I and II.
Conclusions: The RFA system contributed by Osstell Mentor® renders almost perfect reproducibility and repeatability, as proven by statistical analysis carried out by means of ICC with 95% confidence level. This instrument contributes highly reliable RFA measurements in dental implants.
Figures
Similar articles
-
Assessment of Osstell ISQ's reliability for implant stability measurement: a cross-sectional clinical study.Med Oral Patol Oral Cir Bucal. 2013 Nov 1;18(6):e877-82. doi: 10.4317/medoral.19120. Med Oral Patol Oral Cir Bucal. 2013. PMID: 24121909 Free PMC article.
-
Comparative analysis of 2 resonance frequency measurement devices: Osstell Mentor and Osstell ISQ.Implant Dent. 2014 Jun;23(3):351-6. doi: 10.1097/ID.0000000000000072. Implant Dent. 2014. PMID: 24776942
-
Intra- and inter-operator concordance of the resonance frequency analysis. A cross-sectional and prospective clinical study.Clin Oral Investig. 2022 Nov;26(11):6521-6530. doi: 10.1007/s00784-022-04601-y. Epub 2022 Jul 8. Clin Oral Investig. 2022. PMID: 35804172
-
Resonance frequency analysis: Comparing two clinical instruments.Clin Implant Dent Relat Res. 2018 Jun;20(3):308-312. doi: 10.1111/cid.12598. Epub 2018 Mar 13. Clin Implant Dent Relat Res. 2018. PMID: 29532586
-
In vitro comparative analysis of two resonance frequency measurement devices: Osstell implant stability coefficient and Penguin resonance frequency analysis.Clin Implant Dent Relat Res. 2019 Dec;21(6):1124-1131. doi: 10.1111/cid.12868. Epub 2019 Nov 12. Clin Implant Dent Relat Res. 2019. PMID: 31715656
Cited by
-
Clinical evaluation of osseointegration using resonance frequency analysis.J Indian Prosthodont Soc. 2015 Jul-Sep;15(3):192-9. doi: 10.4103/0972-4052.165171. J Indian Prosthodont Soc. 2015. PMID: 26929512 Free PMC article. Review.
-
Effects of Different Undersizing Site Preparations on Implant Stability.Int J Environ Res Public Health. 2020 Dec 2;17(23):8965. doi: 10.3390/ijerph17238965. Int J Environ Res Public Health. 2020. PMID: 33276529 Free PMC article.
-
Stability of dental implants with sandblasted and acid-etched (SLA) and modified (SLActive) surfaces during the osseointegration period.J Dent Res Dent Clin Dent Prospects. 2021 Fall;15(4):226-231. doi: 10.34172/joddd.2021.037. Epub 2021 Dec 5. J Dent Res Dent Clin Dent Prospects. 2021. PMID: 35070174 Free PMC article.
-
Reliability of the Resonance Frequency Analysis Values in New Prototype Transepithelial Abutments: A Prospective Clinical Study.Int J Environ Res Public Health. 2020 Sep 16;17(18):6733. doi: 10.3390/ijerph17186733. Int J Environ Res Public Health. 2020. PMID: 32947802 Free PMC article.
-
Stability of Dental Implants and Thickness of Cortical Bone: Clinical Research and Future Perspectives. A Systematic Review.Materials (Basel). 2021 Nov 25;14(23):7183. doi: 10.3390/ma14237183. Materials (Basel). 2021. PMID: 34885335 Free PMC article. Review.
References
-
- Buser D, Mericske-Stern R, Bernard JP, Behneke A, Behneke N, Hirt HP et al. Long-term evaluation of nonsubmerged ITI implants. Part 1: 8-year life table analysis of a prospective multicenter study with 2359 implants. Clin Oral Implants Res. 1997;8:161–172. - PubMed
-
- Gapski R, Wang HL, Mascarenhas P, Lang NP. Critical review of immediate implant loading. Clin Oral Implants Res. 2003;14:515–27. - PubMed
-
- Rocci A, Martignoni M, Burgos PM, Gottlow J, Sennerby L. Histology of retrieved immediately and early loaded oxidizedimplants:Light microscopic observations after 5 to 9 monthsof loading in the posterior mandible. Clin Implant Dent Relat Res. 2003;5(suppl 1):88–98. - PubMed
-
- Degidi M, Scarano A, Petrone G, Piattelli A. Histological analysis of clinically retrieved immediately loaded implants:Areport of 11 cases. Clin Implant Dent Relat Res. 2003;5:89–93. - PubMed
-
- Weber HP, Morton D, Gallucci GO, Roccuzzo M, Cordaro L, Grutter L. Consensus statements and recommended clinical procedures regarding loading protocols. Int J Oral Maxillofac Implants. 2009;24 Suppl:180–3. - PubMed
MeSH terms
Substances
LinkOut - more resources
Full Text Sources