Safety and accuracy of a new long-term subconjunctival glucose sensor
- PMID: 22341132
- DOI: 10.1111/j.1753-0407.2012.00192.x
Safety and accuracy of a new long-term subconjunctival glucose sensor
Abstract
Background: A new biosensor has been developed by EyeSense (Großostheim, Germany) that is placed into the conjunctiva of one eye to measure the glucose concentration of the surrounding tissue in a non-invasive manner. In the present study we investigated the correlation between glucose concentrations measured by the EyeSense implant and those determined by finger prick testing, as well as the tolerability and safety of the implant over a 16-week period.
Methods: The study was performed in 28 diabetic patients. The biosensor was inserted under local anesthesia and sterile conditions. Correlations between capillary glucose measured by laboratory methods and interstitial glucose determined by the biosensor were investigated by inducing increases and decreases in glucose values between 60 and 300 mg/dL.
Results: Most patients experienced a mild subconjunctival hemorrhage postoperatively. Except for the minor sensation of the presence of foreign body, the implants were well tolerated. Three patients lost the ocular mini insert spontaneously, whereas there was a function failure of the insert in four patients. Error grid analysis showed that the percentage of data pairs in the acceptable ranges (zone A and B) was very high (>96%). However, there was a shift from zone A to zone B during observation. This was due primarily to an increase in the lag time between capillary and interstitial measured glucose.
Conclusion: he present study demonstrates good tolerability and measurement performance of the biosensor. The reasons for an increase in the lag time are still unknown; local reactions may be involved.
Trial registration: ClinicalTrials.gov NCT00999856.
© 2012 Ruijin Hospital, Shanghai Jiaotong University School of Medicine and Wiley Publishing Asia Pty Ltd.
Similar articles
-
Investigation on the accuracy of the blood glucose monitoring device Prestige IQ.Diabetes Nutr Metab. 2003 Aug;16(4):257-61. Diabetes Nutr Metab. 2003. PMID: 14768776 Clinical Trial.
-
Non-invasive glucose monitoring in patients with diabetes: a novel system based on impedance spectroscopy.Biosens Bioelectron. 2006 Dec 15;22(5):598-604. doi: 10.1016/j.bios.2006.01.031. Epub 2006 Mar 9. Biosens Bioelectron. 2006. PMID: 16524714
-
[Current portable blood sugar analysis equipment--comparison of the MediSense Card, Glucometer Elite, Hypocount Supreme and Omnican Control with the glucose oxidase reference method].Schweiz Med Wochenschr. 1997 Jan 7;127(1-2):12-7. Schweiz Med Wochenschr. 1997. PMID: 9036524 German.
-
[Subcutaneously implantable glucose sensors in patients with diabetes mellitus; still many problems].Ned Tijdschr Geneeskd. 2002 Jul 13;146(28):1313-6. Ned Tijdschr Geneeskd. 2002. PMID: 12148218 Review. Dutch.
-
Is it possible to constantly and accurately monitor blood sugar levels, in people with Type 1 diabetes, with a discrete device (non-invasive or invasive)?Diabet Med. 2020 Apr;37(4):532-544. doi: 10.1111/dme.13942. Epub 2019 Mar 13. Diabet Med. 2020. PMID: 30803028 Review.
Cited by
-
Ocular Complications of Diabetes and Therapeutic Approaches.Biomed Res Int. 2016;2016:3801570. doi: 10.1155/2016/3801570. Epub 2016 Mar 28. Biomed Res Int. 2016. PMID: 27119078 Free PMC article. Review.
-
Analytical performance of glucose monitoring systems at different blood glucose ranges and analysis of outliers in a clinical setting.J Diabetes Sci Technol. 2014 May;8(3):466-72. doi: 10.1177/1932296814522804. Epub 2014 Feb 18. J Diabetes Sci Technol. 2014. PMID: 24876607 Free PMC article.
-
Blood glucose self-monitoring with a long-term subconjunctival glucose sensor.J Diabetes Sci Technol. 2013 Jan 1;7(1):24-34. doi: 10.1177/193229681300700104. J Diabetes Sci Technol. 2013. PMID: 23439157 Free PMC article. Clinical Trial.
-
First clinical evaluation of a new percutaneous optical fiber glucose sensor for continuous glucose monitoring in diabetes.J Diabetes Sci Technol. 2013 Jan 1;7(1):13-23. doi: 10.1177/193229681300700103. J Diabetes Sci Technol. 2013. PMID: 23439156 Free PMC article. Clinical Trial.
Publication types
MeSH terms
Substances
Associated data
LinkOut - more resources
Full Text Sources
Medical