First-in-human testing of a wirelessly controlled drug delivery microchip
- PMID: 22344516
- DOI: 10.1126/scitranslmed.3003276
First-in-human testing of a wirelessly controlled drug delivery microchip
Abstract
The first clinical trial of an implantable microchip-based drug delivery device is discussed. Human parathyroid hormone fragment (1-34) [hPTH(1-34)] was delivered from the device in vivo. hPTH(1-34) is the only approved anabolic osteoporosis treatment, but requires daily injections, making patient compliance an obstacle to effective treatment. Furthermore, a net increase in bone mineral density requires intermittent or pulsatile hPTH(1-34) delivery, a challenge for implantable drug delivery products. The microchip-based devices, containing discrete doses of lyophilized hPTH(1-34), were implanted in eight osteoporotic postmenopausal women for 4 months and wirelessly programmed to release doses from the device once daily for up to 20 days. A computer-based programmer, operating in the Medical Implant Communications Service band, established a bidirectional wireless communication link with the implant to program the dosing schedule and receive implant status confirming proper operation. Each woman subsequently received hPTH(1-34) injections in escalating doses. The pharmacokinetics, safety, tolerability, and bioequivalence of hPTH(1-34) were assessed. Device dosing produced similar pharmacokinetics to multiple injections and had lower coefficients of variation. Bone marker evaluation indicated that daily release from the device increased bone formation. There were no toxic or adverse events due to the device or drug, and patients stated that the implant did not affect quality of life.
Comment in
-
Reengineering device translation timelines.Sci Transl Med. 2012 Feb 22;4(122):122ed1. doi: 10.1126/scitranslmed.3003687. Epub 2012 Feb 16. Sci Transl Med. 2012. PMID: 22344517 No abstract available.
-
Microelectronic control of drug delivery.Expert Rev Med Devices. 2012 Jul;9(4):323-6. doi: 10.1586/erd.12.34. Expert Rev Med Devices. 2012. PMID: 22905837
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources