. 2012 Jul;250(7):1023-7.
doi: 10.1007/s00417-012-1953-4. Epub 2012 Feb 21.

Visualization of irrigation fluid flow and calculation of its velocity distribution in the anterior chamber by particle image velocimetry

Affiliations

Visualization of irrigation fluid flow and calculation of its velocity distribution in the anterior chamber by particle image velocimetry

Yuichi Kaji et al. Graefes Arch Clin Exp Ophthalmol. 2012 Jul.

Abstract

Purpose: To visualize irrigation fluid flow and calculate its velocity distribution in the anterior chamber during phacoemulsification by particle image velocimetry.

Methods: Porcine eyes were fixed in a glass chamber filled with balanced salt solution. An ultrasound handpiece was fixed to the glass chamber, and its tip was inserted into the anterior chamber through a corneal incision. Irrigation fluid was mixed with fluorescein-labeled liposomes as tracer particles. During phacoemulsification without ultrasound, a sheet-like Nd-YAG pulsed laser beam was emitted and moved from the iris plane to the top of the cornea continuously. Images of illuminated liposomes in the anterior chamber were captured at short intervals with a CCD camera, and the velocity distribution of irrigation fluid flow was calculated by particle image velocimetry.

Results: By particle image velocimetry, the flow velocity distribution could be calculated in any plane of the anterior chamber. Dynamic flow of the irrigation fluid, ejected from the tip of the ultrasound handpiece and returned to an aspiration port, was visualized clearly in the anterior chamber. The maximum flow velocity in the anterior chamber was 342 ± 131 mm/s.

Conclusions: Particle image velocimetry enabled the visualization of irrigation fluid flow and quantification of its velocity distribution in different planes of the anterior chamber during cataract surgery. These data are essential for evaluating the safety and efficacy of new surgical settings and devices during phacoemulsification.

PubMed Disclaimer

Similar articles

Cited by

References

    1. J Cataract Refract Surg. 2005 Feb;31(2):406-11 - PubMed
    1. J Cataract Refract Surg. 2004 Aug;30(8):1759-67 - PubMed
    1. Lab Chip. 2009 Sep 7;9(17):2551-67 - PubMed
    1. Cornea. 2005 Nov;24(8 Suppl):S55-S58 - PubMed
    1. Biomed Microdevices. 2010 Oct;12(5):875-86 - PubMed

Publication types

LinkOut - more resources