Normal and altered three-dimensional portal venous hemodynamics in patients with liver cirrhosis
- PMID: 22357888
- DOI: 10.1148/radiol.11110127
Normal and altered three-dimensional portal venous hemodynamics in patients with liver cirrhosis
Abstract
Purpose: To compare time-resolved three-dimensional (3D) phase-contrast magnetic resonance (MR) imaging with three-directional velocity encoding (flow-sensitive four-dimensional [4D] MR imaging), with Doppler ultrasonography (US) as standard of reference, for investigating alterations in 3D portal venous hemodynamics in patients with liver cirrhosis compared with healthy age-matched control subjects and healthy young volunteers.
Material & methods: This prospective study was approved by the local ethics committee, and written informed consent was obtained from all participants. Three-dimensional portal venous hemodynamics was assessed, employing flow-sensitive 4D MR imaging with a 3-T MR system (spatial resolution, approximately 2 mm(3); temporal resolution, approximately 45 msec) in 20 patients with hepatic cirrhosis, 20 healthy age-matched control subjects, and 21 healthy young volunteers. Flow characteristics were analyzed by using 3D streamlines and time-resolved particle traces. Quantitative analyses were performed by retrospectively evaluating regional peak and mean velocities, flow volume, and vessel area. Doppler US was used as standard of reference. Independent-sample t tests or Wilcoxon-Mann-Whitney tests were applied for comparing each subject group. Paired-sample t tests or Wilcoxon tests were applied when comparing MR imaging and US.
Results: Three-dimensional visualization of portal venous hemodynamics was successful, with complete visualization of the vessels in 18 patients and 35 volunteers, with limitations in the left intrahepatic branches (87%, reader A; 89%, reader B). A moderate but significant correlation was observed between 4D MR imaging and Doppler US in nearly all maximum and mean velocities, flow volumes, and vessel areas (r = 0.24-0.64, P = .001-.044). With MR imaging, significant underestimation was observed of intrahepatic flow velocities and flow volumes, except vessel area, which Doppler US represented as even lower (P < .001 to P = .045). Six patients had collateralization with reopened umbilical vein, while one had flow reversal in the superior mesenteric vein visible at MR imaging only.
Conclusion: Flow-sensitive 4D MR imaging may constitute a promising, alternative technique to Doppler US for evaluating hemodynamics in the portal venous system of patients with liver cirrhosis and may be a means of assessing pathologic changes in flow characteristics.
© RSNA, 2012.
Similar articles
-
A feasibility study to evaluate splanchnic arterial and venous hemodynamics by flow-sensitive 4D MRI compared with Doppler ultrasound in patients with cirrhosis and controls.Eur J Gastroenterol Hepatol. 2013 Jun;25(6):669-75. doi: 10.1097/MEG.0b013e32835e1297. Eur J Gastroenterol Hepatol. 2013. PMID: 23411868
-
Hepatic flow parameters measured with MR imaging and Doppler US: correlations with degree of cirrhosis and portal hypertension.Radiology. 2003 Nov;229(2):409-14. doi: 10.1148/radiol.2292021128. Epub 2003 Sep 11. Radiology. 2003. PMID: 12970464
-
Quantitative Doppler evaluation of the splenoportal venous system in various stages of cirrhosis: differences between right and left portal veins.J Clin Ultrasound. 2002 Nov-Dec;30(9):537-43. doi: 10.1002/jcu.10114. J Clin Ultrasound. 2002. PMID: 12404519
-
Clinical Applications of 4D Flow MRI in the Portal Venous System.Magn Reson Med Sci. 2022 Mar 1;21(2):340-353. doi: 10.2463/mrms.rev.2021-0105. Epub 2022 Jan 25. Magn Reson Med Sci. 2022. PMID: 35082218 Free PMC article. Review.
-
Four-dimensional flow magnetic resonance imaging in cirrhosis.World J Gastroenterol. 2016 Jan 7;22(1):89-102. doi: 10.3748/wjg.v22.i1.89. World J Gastroenterol. 2016. PMID: 26755862 Free PMC article. Review.
Cited by
-
4D flow imaging: current status to future clinical applications.Curr Cardiol Rep. 2014 May;16(5):481. doi: 10.1007/s11886-014-0481-8. Curr Cardiol Rep. 2014. PMID: 24700368 Review.
-
Reproducibility study of four-dimensional flow MRI of arterial and portal venous liver hemodynamics: influence of spatio-temporal resolution.Magn Reson Med. 2014 Aug;72(2):477-84. doi: 10.1002/mrm.24939. Epub 2013 Sep 9. Magn Reson Med. 2014. PMID: 24018798 Free PMC article.
-
Comparison of blood flow velocity quantification by 4D flow MR imaging with ultrasound at the carotid bifurcation.AJNR Am J Neuroradiol. 2013 Jul;34(7):1407-13. doi: 10.3174/ajnr.A3419. Epub 2013 Feb 14. AJNR Am J Neuroradiol. 2013. PMID: 23413247 Free PMC article.
-
Beneficial long term effect of a phosphodiesterase-5-inhibitor in cirrhotic portal hypertension: A case report with 8 years follow-up.World J Gastroenterol. 2018 Jan 21;24(3):438-444. doi: 10.3748/wjg.v24.i3.438. World J Gastroenterol. 2018. PMID: 29391766 Free PMC article.
-
Splanchnic hemodynamic parameters measured with 4D flow MRI can diagnose severe portal hypertension.Eur Radiol. 2025 Jul 15. doi: 10.1007/s00330-025-11799-4. Online ahead of print. Eur Radiol. 2025. PMID: 40664866
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical