Ultrasonic scalpel causes greater depth of soft tissue necrosis compared to monopolar electrocautery at standard power level settings in a pig model
- PMID: 22361346
- PMCID: PMC3305372
- DOI: 10.1186/1471-2482-12-3
Ultrasonic scalpel causes greater depth of soft tissue necrosis compared to monopolar electrocautery at standard power level settings in a pig model
Abstract
Background: Ultrasonic scalpel (UC) and monopolar electrocautery (ME) are common tools for soft tissue dissection. However, morphological data on the related tissue alteration are discordant. We developed an automatic device for standardized sample excision and compared quality and depth of morphological changes caused by UC and ME in a pig model.
Methods: 100 tissue samples (5 × 3 cm) of the abdominal wall were excised in 16 pigs. Excisions were randomly performed manually or by using the self-constructed automatic device at standard power levels (60 W cutting in ME, level 5 in UC) for abdominal surgery. Quality of tissue alteration and depth of coagulation necrosis were examined histopathologically. Device (UC vs. ME) and mode (manually vs. automatic) effects were studied by two-way analysis of variance at a significance level of 5%.
Results: At the investigated power level settings UC and ME induced qualitatively similar coagulation necroses. Mean depth of necrosis was 450.4 ± 457.8 μm for manual UC and 553.5 ± 326.9 μm for automatic UC versus 149.0 ± 74.3 μm for manual ME and 257.6 ± 119.4 μm for automatic ME. Coagulation necrosis was significantly deeper (p < 0.01) when UC was used compared to ME. The mode of excision (manual versus automatic) did not influence the depth of necrosis (p = 0.85). There was no significant interaction between dissection tool and mode of excision (p = 0.93).
Conclusions: Thermal injury caused by UC and ME results in qualitatively similar coagulation necrosis. The depth of necrosis is significantly greater in UC compared to ME at investigated standard power levels.
Figures





Similar articles
-
Comparison of the effects of surgical dissection devices on the rabbit liver.Surg Today. 2014 Jun;44(6):1116-22. doi: 10.1007/s00595-013-0712-4. Epub 2013 Sep 5. Surg Today. 2014. PMID: 24006128
-
Use of ultrasonic scalpel and monopolar electrocautery for skin incisions in neck dissection: a prospective randomized trial.Oral Maxillofac Surg. 2018 Jun;22(2):169-175. doi: 10.1007/s10006-018-0686-x. Epub 2018 Feb 28. Oral Maxillofac Surg. 2018. PMID: 29492789 Clinical Trial.
-
A single-blind controlled study of electrocautery and ultrasonic scalpel smoke plumes in laparoscopic surgery.Surg Endosc. 2012 Feb;26(2):337-42. doi: 10.1007/s00464-011-1872-1. Epub 2011 Sep 5. Surg Endosc. 2012. PMID: 21898022 Clinical Trial.
-
Ultrasonic dissection versus conventional electrocautery during gastrectomy for gastric cancer: a meta-analysis of randomized controlled trials.Eur J Surg Oncol. 2015 Apr;41(4):527-33. doi: 10.1016/j.ejso.2015.01.025. Epub 2015 Feb 4. Eur J Surg Oncol. 2015. PMID: 25690648 Review.
-
The outcome of laparoscopic cholecystectomy by ultrasonic dissection.Surg Technol Int. 2010 Apr;19:70-8. Surg Technol Int. 2010. PMID: 20437348 Review.
Cited by
-
Evaluation of emissivity and temperature profile of laparoscopic ultrasonic devices (blades and passive jaws).Surg Endosc. 2015 May;29(5):1179-84. doi: 10.1007/s00464-014-3787-0. Epub 2014 Aug 27. Surg Endosc. 2015. PMID: 25159635
-
Comparison of the effects of surgical dissection devices on the rabbit liver.Surg Today. 2014 Jun;44(6):1116-22. doi: 10.1007/s00595-013-0712-4. Epub 2013 Sep 5. Surg Today. 2014. PMID: 24006128
-
Use of ultrasonic scalpel and monopolar electrocautery for skin incisions in neck dissection: a prospective randomized trial.Oral Maxillofac Surg. 2018 Jun;22(2):169-175. doi: 10.1007/s10006-018-0686-x. Epub 2018 Feb 28. Oral Maxillofac Surg. 2018. PMID: 29492789 Clinical Trial.
-
Long-term outcomes of ultrasonic scalpel treatment in giant cell tumor of long bones.Oncol Lett. 2014 Jul;8(1):145-150. doi: 10.3892/ol.2014.2092. Epub 2014 Apr 25. Oncol Lett. 2014. PMID: 24959235 Free PMC article.
-
Bigger Is Not Always Better: Effects of Electrocautery Setting on Tissue Injury in a Porcine Model.Cureus. 2022 Jul 14;14(7):e26841. doi: 10.7759/cureus.26841. eCollection 2022 Jul. Cureus. 2022. PMID: 35974853 Free PMC article.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources