Comparative Study
. 2012 Feb 28;59(9):813-8.
doi: 10.1016/j.jacc.2011.10.893.

Bioprosthetic tricuspid valve regurgitation associated with pacemaker or defibrillator lead implantation

Affiliations
Free article
Comparative Study

Bioprosthetic tricuspid valve regurgitation associated with pacemaker or defibrillator lead implantation

Mackram F Eleid et al. J Am Coll Cardiol. .
Free article

Abstract

Objectives: The goal of this study was to determine the impact of transvenous pacemaker and defibrillator leads on the incidence of bioprosthetic tricuspid valve (BTV) regurgitation compared with BTV patients without a transvalvular lead.

Background: Although concern has been raised regarding the potential deleterious effect of permanent transvenous device leads on BTV function, little is known about the incidence of prosthetic tricuspid regurgitation (TR) after lead placement.

Methods: A retrospective review of 58 patients who underwent BTV implantation and subsequently required endocardial pacemaker (n = 52) or defibrillator (n = 6) lead implantation across the BTV was conducted. Patient and prosthesis characteristics, lead type, and clinical events were collected. The incidence and severity of prosthetic TR, determined by Doppler echocardiography, was compared with 265 consecutive patients who underwent BTV implantation without undergoing subsequent transvalvular device lead implantation.

Results: Over a mean follow-up of 25 months, in 5 patients (9%) with a transvalvular lead significant (moderate or greater) prosthetic TR developed compared with 12 patients (5%) in the control group (p = 0.20). Kaplan-Meier analysis revealed no significant difference in the incidence of TR in BTV patients with and without transvalvular leads (p = 0.45). Significant prosthetic TR in patients with and without a transvalvular lead more commonly occurred 2 years or later after lead or BTV implantation (4 of 5, 80% and 10 of 12, 83%, respectively).

Conclusions: Transvalvular device lead implantation in BTV patients was not associated with an increased incidence of significant prosthetic TR (p = 0.45). Based on these data, transvalvular lead implantation appears to be an acceptable approach for patients with a BTV who require permanent pacemaker or defibrillator placement.