3-dimensional aortic annular assessment by multidetector computed tomography predicts moderate or severe paravalvular regurgitation after transcatheter aortic valve replacement: a multicenter retrospective analysis
- PMID: 22365423
- DOI: 10.1016/j.jacc.2011.12.015
3-dimensional aortic annular assessment by multidetector computed tomography predicts moderate or severe paravalvular regurgitation after transcatheter aortic valve replacement: a multicenter retrospective analysis
Abstract
Objectives: This study sought to analyze multidetector computed tomography (MDCT) 3-dimensional aortic annular dimensions for the prediction of paravalvular aortic regurgitation (PAR) following transcatheter aortic valve replacement (TAVR).
Background: Moderate or severe PAR after TAVR is associated with increased morbidity and mortality.
Methods: A total of 109 consecutive patients underwent MDCT pre-TAVR with a balloon expandable aortic valve. Differences between transcatheter heart valve (THV) size and MDCT measures of annular size (mean diameter, area, and circumference) were analyzed concerning prediction of PAR. Patients with THV malposition (n = 7) were excluded. In 50 patients, MDCT was repeated after TAVR to assess THV eccentricity (1 - short diameter/long diameter) and expansion (MDCT measured THV area/nominal THV area).
Results: Moderate or severe PAR (13 of 102) was associated with THV undersizing (THV diameter - mean diameter = -0.7 ± 1.4 mm vs. 0.9 ± 1.8 mm for trivial to mild PAR, p < 0.01). The difference between THV size and MDCT annular size was predictive of PAR (mean diameter: area under the curve [AUC]: 0.81, 95% confidence interval [CI]: 0.68 to 0.88; area: AUC: 0.80, 95% CI: 0.65 to 0.90; circumference: AUC: 0.76, 95% CI: 0.59 to 0.91). Annular eccentricity was not associated with PAR (AUC: 0.58, 95% CI: 0.46 to 0.75). We found that 35.3% (36 of 102) and 45.1% (46 of 102) of THVs were undersized relative to the MDCT mean diameter and area, respectively. THV oversizing relative to the annular area was not associated with THV eccentricity or underexpansion (oversized vs. undersized THVs; expansion: 102.7 ± 5.3% vs. 106.1 ± 5.6%, p = 0.03; eccentricity: median: 1.7% [interquartile range: 1.4% to 3.0%] vs. 1.7% [interquartile range: 1.1% to 2.7%], p = 0.28).
Conclusions: MDCT-derived 3-dimensional aortic annular measurements are predictive of moderate or severe PAR following TAVR. Oversizing of THVs may reduce the risk of moderate or severe PAR.
Copyright © 2012 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.
Comment in
-
Valvular disease. Super size me? Annular size and paravalvular leak after TAVR.Nat Rev Cardiol. 2012 Mar 20;9(5):265-6. doi: 10.1038/nrcardio.2012.38. Nat Rev Cardiol. 2012. PMID: 22430832
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical