Dosimetric impact of breathing motion in lung stereotactic body radiotherapy treatment using intensity modulated radiotherapy and volumetric modulated arc therapy [corrected]
- PMID: 22365622
- DOI: 10.1016/j.ijrobp.2011.12.001
Dosimetric impact of breathing motion in lung stereotactic body radiotherapy treatment using intensity modulated radiotherapy and volumetric modulated arc therapy [corrected]
Erratum in
- Int J Radiat Oncol Biol Phys. 2013 Dec 1;87(5):859
- Int J Radiat Oncol Biol Phys. 2014 Feb 1;88(2):462
Abstract
Purpose: The objective of this study was to investigate the influence of tumor motion on dose delivery in stereotactic body radiotherapy (SBRT) for lung cancer, using fixed field intensity- modulated radiotherapy (IMRT) and volumetric modulated arc therapy (VMAT).
Methods and materials: For each of 10 patients with stage I/II non-small-cell pulmonary tumors, a respiration-correlated four-dimensional computed tomography (4DCT) study was carried out. The internal target volume was delineated on the maximum intensity projection CT, which was reconstructed from the 4DCT dataset. A 5-mm margin was used for generation of the planning target volume. VMAT and five-field IMRT plans were generated using Pinnacle(3) SmartArc and direct machine parameter optimization, respectively. All plans were generated for an Elekta Synergy linear accelerator using 6-MV photons. Simulation was performed to study the interplay between multileaf collimator (MLC) sequences and target movement during the delivery of VMAT and IMRT. For each plan, 4D dose was calculated using deformable image registration of the 4DCT images. Target volume coverage and doses to critical structures calculated using 4D methodology were compared with those calculated using 3D methodology.
Results: For all patients included in this study, the interplay effect was found to present limited impact (less than 1% of prescription) on the target dose distribution, especially for SBRT, in which fewer fractions (three fractions) are delivered. Dose to the gross tumor volume (GTV) was, on average, slightly decreased (1% of prescription) in the 4D calculation compared with the 3D calculation. The motion impact on target dose homogeneity was patient-dependent and relatively small.
Conclusions: Both VMAT and IMRT plans experienced negligible interplay effects between MLC sequence and tumor motion. For the most part, the 3D doses to the GTV and critical structures provided good approximations of the 4D dose calculations.
Copyright © 2012 Elsevier Inc. All rights reserved.
Similar articles
-
4D VMAT, gated VMAT, and 3D VMAT for stereotactic body radiation therapy in lung.Phys Med Biol. 2013 Feb 21;58(4):749-70. doi: 10.1088/0031-9155/58/4/749. Epub 2013 Jan 16. Phys Med Biol. 2013. PMID: 23324560
-
On the validity of density overrides for VMAT lung SBRT planning.Med Phys. 2014 Aug;41(8):081707. doi: 10.1118/1.4887778. Med Phys. 2014. PMID: 25086517
-
Assessment of Monte Carlo algorithm for compliance with RTOG 0915 dosimetric criteria in peripheral lung cancer patients treated with stereotactic body radiotherapy.J Appl Clin Med Phys. 2016 May 8;17(3):277-293. doi: 10.1120/jacmp.v17i3.6077. J Appl Clin Med Phys. 2016. PMID: 27167284 Free PMC article.
-
Influence of the jaw tracking technique on the dose calculation accuracy of small field VMAT plans.J Appl Clin Med Phys. 2017 Jan;18(1):186-195. doi: 10.1002/acm2.12029. Epub 2017 Jan 3. J Appl Clin Med Phys. 2017. PMID: 28291941 Free PMC article. Review.
-
[Intensity-modulated radiation therapy in non-small cell lung cancers].Cancer Radiother. 2014 Oct;18(5-6):406-13. doi: 10.1016/j.canrad.2014.06.009. Epub 2014 Sep 16. Cancer Radiother. 2014. PMID: 25219608 Review. French.
Cited by
-
A method of dose reconstruction for moving targets compatible with dynamic treatments.Med Phys. 2012 Oct;39(10):6237-46. doi: 10.1118/1.4754297. Med Phys. 2012. PMID: 23039659 Free PMC article.
-
Dose deviations induced by respiratory motion for radiotherapy of lung tumors: Impact of CT reconstruction, plan complexity, and fraction size.J Appl Clin Med Phys. 2020 Apr;21(4):68-79. doi: 10.1002/acm2.12847. Epub 2020 Mar 12. J Appl Clin Med Phys. 2020. PMID: 32166850 Free PMC article.
-
The Practicality of ICRU and Considerations for Future ICRU Definitions.Semin Radiat Oncol. 2018 Jun;28(3):201-206. doi: 10.1016/j.semradonc.2018.02.005. Semin Radiat Oncol. 2018. PMID: 29933880 Free PMC article. Review.
-
Evaluation of the target dose coverage of stereotactic body radiotherapy for lung cancer using helical tomotherapy: A dynamic phantom study.Rep Pract Oncol Radiother. 2020 Mar-Apr;25(2):200-205. doi: 10.1016/j.rpor.2020.01.001. Epub 2020 Jan 14. Rep Pract Oncol Radiother. 2020. PMID: 32021577 Free PMC article.
-
Influence of segment width on plan quality for volumetric modulated arc based stereotactic body radiotherapy.Rep Pract Oncol Radiother. 2014 Apr 16;19(5):287-95. doi: 10.1016/j.rpor.2014.03.003. eCollection 2014 Sep. Rep Pract Oncol Radiother. 2014. PMID: 25184052 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical