Accuracy of linear measurements in cone beam computed tomography with different voxel sizes
- PMID: 22382754
- DOI: 10.1097/ID.0b013e31824bf93c
Accuracy of linear measurements in cone beam computed tomography with different voxel sizes
Abstract
Objective: To evaluate the accuracy of linear measurements on dry mandible specimens using cone beam computed tomography (CBCT) images acquired with different voxel sizes.
Methodology: Eight human dry mandibles were submitted to CBCT examination, using the i-CAT (Imaging Sciences, Inc. Hatfield, PA) device and four protocols with different voxel sizes. Tomographic slices with a more central view of the markers, placed on six sites on each mandible, were selected to perform measurements. Values obtained from direct measurements on the dry mandible after sectioning them on the preestablished sites were compared with measurements from the tomographic images and the measurement error.
Results: There was no statistical difference between the measurement error of the protocols (P = 0.606). The mean value of the difference between the values obtained in the images and the dry mandible was smaller than 1 mm for all the protocols.
Conclusion: The accuracy of vertical and horizontal measurements, using CBCT (i-CAT) for the four protocols, was shown to be comparable with the measurements performed on the dry mandible.
Similar articles
-
Accuracy and reliability of buccal bone height and thickness measurements from cone-beam computed tomography imaging.Am J Orthod Dentofacial Orthop. 2011 Nov;140(5):734-44. doi: 10.1016/j.ajodo.2011.06.021. Am J Orthod Dentofacial Orthop. 2011. PMID: 22051495
-
Influence of scan mode (180°/360°) of the cone beam computed tomography for preoperative dental implant measurements.Clin Oral Implants Res. 2014 Feb;25(2):e155-8. doi: 10.1111/clr.12080. Epub 2012 Dec 12. Clin Oral Implants Res. 2014. PMID: 23231370
-
Accuracy of bone surface size and cortical layer thickness measurements using cone beam computerized tomography.Clin Oral Implants Res. 2013 Jul;24(7):793-7. doi: 10.1111/j.1600-0501.2012.02459.x. Epub 2012 Apr 2. Clin Oral Implants Res. 2013. PMID: 22469137
-
Accuracy of linear measurements on CBCT images related to presurgical implant treatment planning: A systematic review.Clin Oral Implants Res. 2018 Oct;29 Suppl 16:393-415. doi: 10.1111/clr.13142. Clin Oral Implants Res. 2018. PMID: 30328204
-
Accuracy of linear measurements for implant planning based on low-dose cone beam CT protocols: a systematic review and meta-analysis.Dentomaxillofac Radiol. 2024 Apr 29;53(4):207-221. doi: 10.1093/dmfr/twae007. Dentomaxillofac Radiol. 2024. PMID: 38429951 Free PMC article.
Cited by
-
Comparison of the influence of FOV sizes and different voxel resolutions for the assessment of periodontal defects.Dentomaxillofac Radiol. 2015;44(7):20150070. doi: 10.1259/dmfr.20150070. Epub 2015 Apr 22. Dentomaxillofac Radiol. 2015. PMID: 25900235 Free PMC article.
-
Comparative study of the accuracy of CBCT implant site measurements using different software programs.Saudi Dent J. 2021 Sep;33(6):355-361. doi: 10.1016/j.sdentj.2020.07.003. Epub 2020 Aug 6. Saudi Dent J. 2021. PMID: 34434039 Free PMC article.
-
Skeletal, Dentoalveolar and Dental Changes after "Mini-Screw Assisted Rapid Palatal Expansion" Evaluated with Cone Beam Computed Tomography.J Clin Med. 2022 Aug 9;11(16):4652. doi: 10.3390/jcm11164652. J Clin Med. 2022. PMID: 36012886 Free PMC article.
-
Mandibular skeletal posterior anatomic limit for molar distalization in patients with Class III malocclusion with different vertical facial patterns.Korean J Orthod. 2021 Jul 25;51(4):250-259. doi: 10.4041/kjod.2021.51.4.250. Korean J Orthod. 2021. PMID: 34275881 Free PMC article.
-
Identification of mandibular canal in cone beam computed tomography plane with different voxel sizes.Saudi Dent J. 2020 Dec;32(8):403-409. doi: 10.1016/j.sdentj.2019.10.009. Epub 2019 Nov 6. Saudi Dent J. 2020. PMID: 33304084 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous