. 2012 Jun 1;83(2):e287-95.
doi: 10.1016/j.ijrobp.2011.12.063. Epub 2012 Mar 6.

An innovative tool for intraoperative electron beam radiotherapy simulation and planning: description and initial evaluation by radiation oncologists

Affiliations
Free article

An innovative tool for intraoperative electron beam radiotherapy simulation and planning: description and initial evaluation by radiation oncologists

Javier Pascau et al. Int J Radiat Oncol Biol Phys. .
Free article

Abstract

Purpose: Intraoperative electron beam radiation therapy (IOERT) involves a modified strategy of conventional radiation therapy and surgery. The lack of specific planning tools limits the spread of this technique. The purpose of the present study is to describe a new simulation and planning tool and its initial evaluation by clinical users.

Methods and materials: The tool works on a preoperative computed tomography scan. A physician contours regions to be treated and protected and simulates applicator positioning, calculating isodoses and the corresponding dose--volume histograms depending on the selected electron energy. Three radiation oncologists evaluated data from 15 IOERT patients, including different tumor locations. Segmentation masks, applicator positions, and treatment parameters were compared.

Results: High parameter agreement was found in the following cases: three breast and three rectal cancer, retroperitoneal sarcoma, and rectal and ovary monotopic recurrences. All radiation oncologists performed similar segmentations of tumors and high-risk areas. The average applicator position difference was 1.2 ± 0.95 cm. The remaining cancer sites showed higher deviations because of differences in the criteria for segmenting high-risk areas (one rectal, one pancreas) and different surgical access simulated (two rectal, one Ewing sarcoma).

Conclusions: The results show that this new tool can be used to simulate IOERT cases involving different anatomic locations, and that preplanning has to be carried out with specialized surgical input.