Rapid visual colorimetry of peritoneal lavage fluid

Affiliations

Rapid visual colorimetry of peritoneal lavage fluid

D Tandberg et al. Ann Emerg Med. 1990 Nov.

Abstract

Study hypothesis: That visual colorimetry can be used to rapidly and precisely estimate the erythrocyte count of 1:5 dilutions of simulated peritoneal lavage fluid.

Population: Fifty-four normal adult human subjects.

Methods: The automated or chamber RBC count is often used on fluid obtained by peritoneal lavage in patients with abdominal trauma to help determine the need for surgery. Unfortunately, this method sometimes results in excessive delay. We designed and built a simple colorimeter that facilitated rapid direct visual comparison of unknown samples with known color standards. A radiograph view box was used as a light source. Standards were prepared in 16-mm glass tubes to simulate peritoneal lavage fluid with RBC counts ranging from 0 to 140,000 in 10,000 cell/microL increments; 1:5 dilutions with water were used throughout to reduce opacity. Thimerosal was added to unknowns and standards to stabilize color; all samples were kept refrigerated at 4 C when not in use. In a double-blind in-vitro study, each subject matched 20 randomly distributed unknowns ranging from 12,000 to 131,000 erythrocytes/microL to the nearest standard.

Results: The mean absolute error for all 1,080 determinations was 3,560 RBC/microL (95% CI = 4,290-4,830; SD = 4,560; t = 39.6; df = 1,079; P less than .001). This method correctly predicted the RBC count to within 9,000 cells/microL 95% of the time.

Conclusion: Visual comparison of 1:5 dilutions of simulated peritoneal lavage fluid with known color standards can be used to rapidly and precisely estimate the erythrocyte count.