Standardization of a 85Sr solution by three methods
- PMID: 22425416
- DOI: 10.1016/j.apradiso.2012.02.103
Standardization of a 85Sr solution by three methods
Abstract
A (85)Sr solution was measured in two liquid scintillation (LS) counters. Scintillation pulses from the delayed transitions in the (85)Sr decay were by and large lost as the dead time of the LS-counter was much longer than the lifetime of the (85)Rb excited level. The registration probability of these pulses has been defined and introduced into the counting efficiency calculation. The results of the (85)Sr solution standardization by the CIEMAT/NIST and the TDCR methods relative to the results from the 4π(LS)-γ coincidence and anticoincidence method were 0.997±0.012 and 0.971±0.019, respectively.
Copyright © 2012 Elsevier Ltd. All rights reserved.