A guide to fluoroscopic identification and design of bioprosthetic valves: a reference for valve-in-valve procedure
- PMID: 22431472
- DOI: 10.1002/ccd.24419
A guide to fluoroscopic identification and design of bioprosthetic valves: a reference for valve-in-valve procedure
Abstract
Surgical aortic valve replacement remains the therapy of choice in majority of patients with aortic stenosis. Bioprosthetic heart valves are often preferred over mechanical valves as they preclude the need for anticoagulation with its associated risks of bleeding and thromboembolism. However, bioprosthetic heart valves undergo structural deterioration and eventually fail. Reoperation is the standard treatment for structural failure of the bioprosthetic valve, stenosis or regurgitation but can carry a significant risk, especially in elderly patients with multiple comorbidities. Transcatheter aortic valve implantation has recently been established as a feasible alternative to conventional valve surgery for the management of high-risk elderly patients with aortic stenosis. This treatment modality has also been shown to be of benefit in the management of degenerated aortic bioprosthesis as a valve-in-valve procedure. The success of this procedure depends on a good understanding of the failing bioprostheses. This not only includes the device design but its radiological/fluoroscopic appearance and how it correlates with the implanted valve, as transcatheter aortic valve implantation is performed under fluoroscopic guidance. Here we illustrate the fluoroscopic appearance of 11 commercially available surgical bioprostheses and two commercially available transcatheter heart valves and discuss important aspects in their design which can influence outcome of a valve-in-valve procedure. We have also collated relevant information on the aspects of the design of a bioprosthetic valve, which are relevant to the valve-in-valve procedure.
Copyright © 2012 Wiley Periodicals, Inc.
Comment in
-
Need to identify bioprosthetic valves: a call for device reconciliation in the electronic medical record.Catheter Cardiovasc Interv. 2013 Apr;81(5):862-3. doi: 10.1002/ccd.24894. Catheter Cardiovasc Interv. 2013. PMID: 23526620 No abstract available.
-
Rebuttal: Fluoroscopic characterization of surgical bioprosthetic heart valves.Catheter Cardiovasc Interv. 2015 Jun;85(7):1277-9. doi: 10.1002/ccd.25902. Catheter Cardiovasc Interv. 2015. PMID: 25712348 No abstract available.
-
Fluoroscopic characterization of surgical bioprosthetic heart valves.Catheter Cardiovasc Interv. 2015 Jun;85(7):1274-6. doi: 10.1002/ccd.25904. Epub 2015 Apr 24. Catheter Cardiovasc Interv. 2015. PMID: 25712584 No abstract available.
Similar articles
-
Transcatheter aortic valve-in-valve implantation for patients with degenerative surgical bioprosthetic valves.Curr Probl Cardiol. 2014 Jan;39(1):7-27. doi: 10.1016/j.cpcardiol.2013.10.001. Epub 2013 Oct 24. Curr Probl Cardiol. 2014. PMID: 24331437 Review.
-
Transcatheter aortic valve implantation in failed bioprosthetic surgical valves.JAMA. 2014 Jul;312(2):162-70. doi: 10.1001/jama.2014.7246. JAMA. 2014. PMID: 25005653
-
Valve-in-valve implantation of Medtronic CoreValve prosthesis in patients with failing bioprosthetic aortic valves.Circ Cardiovasc Interv. 2012 Oct;5(5):689-97. doi: 10.1161/CIRCINTERVENTIONS.112.972331. Epub 2012 Oct 9. Circ Cardiovasc Interv. 2012. PMID: 23048050
-
Transcatheter aortic valve implantation for treatment of patients with degenerated aortic bioprostheses--valve-in-valve technique.Catheter Cardiovasc Interv. 2010 Dec 1;76(7):1000-6. doi: 10.1002/ccd.22738. Epub 2010 Nov 17. Catheter Cardiovasc Interv. 2010. PMID: 20824744
-
Transcatheter Aortic Valve-in-Valve Procedure in Patients with Bioprosthetic Structural Valve Deterioration.Methodist Debakey Cardiovasc J. 2017 Jul-Sep;13(3):132-141. doi: 10.14797/mdcj-13-3-132. Methodist Debakey Cardiovasc J. 2017. PMID: 29743998 Free PMC article. Review.
Cited by
-
Challenges in valve-in-valve therapy.J Thorac Dis. 2015 Sep;7(9):1501-8. doi: 10.3978/j.issn.2072-1439.2015.06.16. J Thorac Dis. 2015. PMID: 26543595 Free PMC article. Review.
-
Transcatheter Aortic Valve Replacement for Degenerative Bioprosthetic Mosaic Valve.JACC Case Rep. 2025 Feb 5;30(3):103124. doi: 10.1016/j.jaccas.2024.103124. eCollection 2025 Feb 5. JACC Case Rep. 2025. PMID: 39963226 Free PMC article.
-
Valve-in-valve implantations: is this the new standard for degenerated bioprostheses? Review of the literature.Clin Res Cardiol. 2014 Jun;103(6):417-29. doi: 10.1007/s00392-013-0653-3. Epub 2014 Jan 21. Clin Res Cardiol. 2014. PMID: 24445751 Review.
-
Radiopaque preventive landmarks' placement during stentless bioprosthesis implantation.J Thorac Dis. 2018 Jun;10(6):3158-3165. doi: 10.21037/jtd.2018.06.16. J Thorac Dis. 2018. PMID: 30069311 Free PMC article.
-
Investigation of failure modes of explanted porcine valves in the mitral position.J Thorac Dis. 2021 May;13(5):2858-2866. doi: 10.21037/jtd-20-3578. J Thorac Dis. 2021. PMID: 34164177 Free PMC article.