Universal quantification of structurally diverse natural products using an evaporative light scattering detector

Affiliations
Free PMC article

Universal quantification of structurally diverse natural products using an evaporative light scattering detector

Navid Adnani et al. J Nat Prod. .
Free PMC article

Abstract

A lack of good methods for absolute quantification of natural products has limited the accuracy of high-throughput screening. Many currently used methods for quantification are either too slow or not amenable to the structural diversity of natural products. Recent developments in low-temperature evaporative light scattering detectors (ELSD-LT) have overcome several historical limitations of ELSDs, including analyte decomposition and low sensitivity. Primarily, ELSDs have been used for relative quantification and detection of compounds that lack a UV chromophore. In this study, we employ an ELSD-LT for absolute quantification of natural products. Calibration curves were constructed using a weighted least-squares analysis for a diverse set of natural products and other compounds. An average calibration curve was evaluated for the "universal" quantification of natural products. Optimization of ELSD-LT hardware and parameters improved sensitivity and throughput and established the utility of ELSD-LT for quantification of large natural product libraries.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Effects of solvent composition and evaporation temperature on ELSD response.
Figure 2
Figure 2
Calibration curve of cycloheximide.
Figure 3
Figure 3
Comparison of standard calibration curves and construction of average “universal” calibration curve.

Similar articles

Cited by

References

    1. Harvey A. L. Drug Discovery Today 2008, 13, 894–901. - PubMed
    1. Feher M.; Schmidt J. M. J. Chem. Inf. Comput. Sci. 2003, 43, 218–227. - PubMed
    1. Li J. W.-H.; Vederas J. C. Science 2009, 325, 161–165. - PubMed
    1. Malo N.; Hanley J. A.; Cerquozzi S.; Pelletier J.; Nadon R. Nat. Biotechnol. 2006, 24, 167–175. - PubMed
    1. Appleton D. R.; Buss A. D.; Butler M. S. Chimia 2007, 61, 327–331.

Publication types

Substances