Simple RP-HPLC method for determination of triple drug combination of valsartan, amlodipine and hydrochlorothiazide in human plasma
- PMID: 22472448
- DOI: 10.2478/v10007-012-0004-3
Simple RP-HPLC method for determination of triple drug combination of valsartan, amlodipine and hydrochlorothiazide in human plasma
Abstract
A simple RP-HPLC method for the quantification of valsartan (VAL), amlodipine (AML) and hydrochlorothiazide (HCT) in human plasma was developed and validated. VAL, AML and HCT were resolved using a Gemini C18 column and mobile phase gradient starting from 20 % acetonitrile and 80 % 10 mmol L(-1) ammonium formate (V/V, pH 3.5 ± 0.2, by formic acid) to 70 % acetonitrile and 30 % 10 mmol L(-1) ammonium formate, over 20 minutes, with a flow rate of 1 mL min(-1). The samples were purified by protein precipitation and extraction. Telmisartan was used as internal standard. The method was validated according to USFDA and EMEA guidelines with good reproducibility and linear responses R = 0.9985 (VAL), 0.9964 (AML), and 0.9971 (HCT). RSDs of intra- and inter-day precision ranged 2.2-8.1 and 4.6-11.7 %, respectively, for all three drugs. Mean extraction recoveries of three QCs for the triple drug combination were 76.5 (VAL), 72.0 (AML) and 73.0 (HCT) % for human plasma. Although the LC-MS/MS method is more sensitive than HPLC, HPLC is still suitable for preliminary pharmacokinetic study. The experiments performed demostrated that simultaneous determination of all components of the triple drug combination in human plasma can be done by this method. Proposed method can be also used for guidance to the LC-MS/MS method.
Similar articles
-
Development and validation of a highly sensitive and robust LC-ESI-MS/MS method for simultaneous quantitation of simvastatin acid, amlodipine and valsartan in human plasma: application to a clinical pharmacokinetic study.Biomed Chromatogr. 2009 Jun;23(6):615-22. doi: 10.1002/bmc.1161. Biomed Chromatogr. 2009. PMID: 19277959
-
A liquid chromatography/tandem mass spectrometry method for the simultaneous quantification of valsartan and hydrochlorothiazide in human plasma.J Chromatogr B Analyt Technol Biomed Life Sci. 2007 Jun 1;852(1-2):436-42. doi: 10.1016/j.jchromb.2007.02.014. Epub 2007 Feb 15. J Chromatogr B Analyt Technol Biomed Life Sci. 2007. PMID: 17331816
-
Reliable and easy-to-use LC-MS/MS-method for simultaneous determination of the antihypertensives metoprolol, amlodipine, canrenone and hydrochlorothiazide in patients with therapy-refractory arterial hypertension.J Pharm Biomed Anal. 2019 Feb 5;164:373-381. doi: 10.1016/j.jpba.2018.11.002. Epub 2018 Nov 5. J Pharm Biomed Anal. 2019. PMID: 30439665
-
Valsartan-amlodipine-hydrochlorothiazide: the definitive fixed combination?Expert Rev Cardiovasc Ther. 2010 Nov;8(11):1609-18. doi: 10.1586/erc.10.115. Expert Rev Cardiovasc Ther. 2010. PMID: 21090936 Review.
-
Amlodipine/valsartan single-pill combination: a review of its use in the management of hypertension.Am J Cardiovasc Drugs. 2009;9(5):309-30. doi: 10.2165/11201120-000000000-00000. Am J Cardiovasc Drugs. 2009. PMID: 19791840 Review.
Cited by
-
Simple simultaneous analysis of various cardiovascular drug mixtures with vincamine: comparative eco-friendly assessment.BMC Chem. 2024 Oct 10;18(1):197. doi: 10.1186/s13065-024-01303-2. BMC Chem. 2024. PMID: 39390600 Free PMC article.
-
Application of an LC-MS/MS method for the analysis of amlodipine, valsartan and hydrochlorothiazide in polypill for a bioequivalence study.J Pharm Anal. 2017 Oct;7(5):309-316. doi: 10.1016/j.jpha.2017.06.001. Epub 2017 Jun 4. J Pharm Anal. 2017. PMID: 29404054 Free PMC article.
-
Development and Validation of a LC-MS/MS Method for the Simultaneous Estimation of Amlodipine and Valsartan in Human Plasma: Application to a Bioequivalence Study.Sci Pharm. 2014 Mar 26;82(3):585-600. doi: 10.3797/scipharm.1402-11. Print 2014 Jul-Sep. Sci Pharm. 2014. PMID: 25853070 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous