Spectral entropy for assessing the depth of propofol sedation
- PMID: 22474549
- PMCID: PMC3315652
- DOI: 10.4097/kjae.2012.62.3.234
Spectral entropy for assessing the depth of propofol sedation
Abstract
Background: For patients in the intensive care unit (ICU) or under monitored anesthetic care (MAC), the precise monitoring of sedation depth facilitates the optimization of dosage and prevents adverse complications from underor over-sedation. For this purpose, conventional subjective sedation scales, such as the Observer's Assessment of Alertness/Sedation (OAA/S) or the Ramsay scale, have been widely utilized. Current procedures frequently disturb the patient's comfort and compromise the already well-established sedation. Therefore, reliable objective sedation scales that do not cause disturbances would be beneficial. We aimed to determine whether spectral entropy can be used as a sedation monitor as well as determine its ability to discriminate all levels of propofol-induced sedation during gradual increments of propofol dosage.
Methods: In 25 healthy volunteers undergoing general anesthesia, the values of response entropy (RE) and state entropy (SE) corresponding to each OAA/S (5 to 1) were determined. The scores were then analyzed during each 0.5 mcg/ml- incremental increase of a propofol dose.
Results: We observed a reduction of both RE and SE values that correlated with the OAA/S (correlation coefficient of 0.819 in RE-OAA/S and 0.753 in SE-OAA/S). The RE and SE values corresponding to awake (OAA/S score 5), light sedation (OAA/S 3-4) and deep sedation (OAA/S 1-2) displayed differences (P < 0.05).
Conclusions: The results indicate that spectral entropy can be utilized as a reliable objective monitor to determine the depth of propofol-induced sedation.
Keywords: Entropy; Propofol; Sedation.
Figures


References
-
- LeBlanc JM, Dasta JF, Kane-Gill SL. Role of the bispectral index in sedation monitoring in the ICU. Ann Pharmacother. 2006;40:490–500. - PubMed
-
- Vivien B, Di Maria S, Ouattara A, Langeron O, Coriat P, Riou B. Overestimation of bispectral index in sedated intensive care unit patients revealed by administration of muscle relaxant. Anesthesiology. 2003;99:9–17. - PubMed
-
- Vanluchene AL, Struys MM, Heyse BE, Mortier EP. Spectral entropy measurement of patient responsiveness during propofol and remifentanil. A comparison with the bispectral index. Br J Anaesth. 2004;93:645–654. - PubMed
-
- Hernández-Gancedo C, Pestaña D, Perez-Chrzanowska H, Martinez-Casanova E, Criado A. Comparing entropy and the bispectral index with the Ramsay score in sedated ICU patients. J Clin Monit Comput. 2007;21:295–302. - PubMed
-
- Schmidt GN, Bischoff P, Standl T, Hellstern A, Teuber O, Schulte Esch J. Comparative evaluation of the Datex-Ohmeda S/5 entropy module and the bispectral index monitor during propofol-remifentanil anesthesia. Anesthesiology. 2004;101:1283–1290. - PubMed
LinkOut - more resources
Full Text Sources