Squeaking hip arthroplasties: a tribological phenomenon
- PMID: 22480525
- DOI: 10.1016/j.arth.2012.01.023
Squeaking hip arthroplasties: a tribological phenomenon
Abstract
The clinical incidence of squeaking has been reported with increasing frequency, with ceramic-on-ceramic bearings seemingly most affected. This study investigated potential causes of squeaking in hard-on-hard hip bearings through 2 sets of experimental conditions. Bearing clearance appeared to affect the incidence of squeaking in metal-on-metal surface arthroplasties. The addition of third-body particles to the interface for total hip arthroplasties also affected the incidence of squeaking. In both studies, the incidence of squeaking correlated well with elevated friction. The findings of this study suggest that a likely cause of squeaking in the hip arthroplasty is adverse tribological conditions caused by suboptimal lubrication. There are numerous factors that may cause the suboptimal lubrication, and therefore, it is unlikely that an individual cause for squeaking will be identified.
Copyright © 2013 Elsevier Inc. All rights reserved.
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical