Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2012 Apr;39(4):2078-89.
doi: 10.1118/1.3694108.

A novel energy mapping approach for CT-based attenuation correction in PET

Affiliations

A novel energy mapping approach for CT-based attenuation correction in PET

B Teimourian et al. Med Phys. 2012 Apr.

Abstract

Purpose: Dual-energy CT (DECT) is arguably the most accurate energy mapping technique in CT-based attenuation correction (CTAC) implemented on hybrid PET/CT systems. However, this approach is not attractive for clinical use owing to increased patient dose. The authors propose a novel energy mapping approach referred to as virtual DECT (VDECT) taking advantage of the DECT formulation but using CT data acquired at a single energy (kV(P)). For this purpose, the CT image acquired at one energy is used to generate the CT image at a second energy using calculated kV(P) conversion curves derived from phantom studies.

Methods: The attenuation map (μ-map) at 511 keV was generated for the XCAT phantom and clinical studies using the bilinear, DECT, and VDECT techniques. The generated μ-maps at 511 keV are compared to the reference derived from the XCAT phantom serving as ground truth. PET data generated from a predefined activity map for the XCAT phantom were then corrected for attenuation using μ-maps generated using the different energy mapping approaches. In addition, the generated μ-maps using the above described methods for a cylindrical polyethylene phantom containing different concentrations of K(2)HPO(4) in water were compared to actual attenuation coefficients. Likewise, CT images of five clinical whole-body studies were used to generate μ-maps using the various energy-mapping approaches were compared with μ-maps acquired at 511 keV using (68)Ge/(68)Ga rod sources for the clinical studies.

Results: The results of phantom studies demonstrate that the proposed method is more accurate than the bilinear technique. All three μ-maps yielded almost similar results for soft and lung tissues whereas for bone tissues, the DECT and the VDECT methods produced a much smaller mean relative difference (3.0% and 2.8%, respectively) than the bilinear approach (11.8%). Likewise, the comparison of PET images corrected for attenuation using the various methods showed that the proposed method provides better accuracy (6.5%) than the bilinear method (13.4%). Clinical studies further demonstrated that, compared to the bilinear method, the VDECT approach has better agreement for bony structures with the DECT technique (1.5% versus 8.9%) and transmission scanning (8.8% versus 17.7%).

Conclusions: It was concluded that the proposed method outperforms the bilinear method especially in bony structures. Further evaluation using a large clinical PET/CT database is underway to evaluate the potential of the technique in a clinical setting.

PubMed Disclaimer

Publication types

LinkOut - more resources