Troponin-specific autoantibody interference in different cardiac troponin I assay configurations
- PMID: 22490617
- DOI: 10.1373/clinchem.2011.179226
Troponin-specific autoantibody interference in different cardiac troponin I assay configurations
Abstract
Background: Autoantibodies to cardiac troponins (cTnAAb) can interfere with the measurement of cardiac troponin I (cTnI) by immunoassays. The aim of this study was to explore the degree of cTnAAb interference in different cTnI assay configurations.
Methods: Ternary troponin complex was added into samples (serum or plasma, n = 132, 68% cTnAAb positive) from individuals without known cardiac conditions. The recovery of cTnI was then measured with 6 investigational cTnI assays (2, 3, or 4 antibodies per assay). Three of these assays were then selected for further comparison by use of samples (plasma, n = 210, 33% cTnAAb positive) from non-ST-elevation acute coronary syndrome patients in the FRISC-II (FRagmin/Fast Revascularisation during InStability in Coronary artery disease) cohort. Finally, these results were compared to those obtained with 3 commercial cTnI assays.
Results: Analytical recoveries varied widely among the 6 investigational assays. Notably the low recoveries (median 9%) of the midfragment-targeting reference assay were normalized (median 103%) with the use of the 4-antibody assay construct (3 capture, 1 tracer antibody) with only 1 antibody against a midfragment epitope. Reduced analytical recoveries correlated closely with measured autoantibody amounts. cTnI concentrations from cTnAAb-positive patient samples determined with 3 investigational assays confirmed the reduced concentrations expected from the low analytical recoveries. The results from the commercial cTnI assays with antibody selections representative for contemporary assay constructs revealed a similar underestimation (up to 20-fold) of cTnI in cTnAAb-positive samples.
Conclusions: A novel cTnI assay deviating from the conventional IFCC-recommended midfragment approach substantially improves cTnI detection in samples containing cTnAAbs.
Similar articles
-
Novel sensitive cardiac troponin I immunoassay free from troponin I-specific autoantibody interference.Clin Chem Lab Med. 2014 Jul;52(7):1041-8. doi: 10.1515/cclm-2013-1044. Clin Chem Lab Med. 2014. PMID: 24615485
-
Epitope specificity and IgG subclass distribution of autoantibodies to cardiac troponin.Clin Chem. 2013 Mar;59(3):512-8. doi: 10.1373/clinchem.2012.194860. Epub 2013 Jan 3. Clin Chem. 2013. PMID: 23288486
-
Comparison of cardiac troponin I immunoassays variably affected by circulating autoantibodies.Clin Chem. 2005 May;51(5):848-55. doi: 10.1373/clinchem.2004.040089. Epub 2005 Feb 17. Clin Chem. 2005. PMID: 15718488
-
Analytical characteristics of high-sensitivity cardiac troponin assays.Clin Chem. 2012 Jan;58(1):54-61. doi: 10.1373/clinchem.2011.165795. Epub 2011 Sep 30. Clin Chem. 2012. PMID: 21965555 Review.
-
Standardization of troponin I measurements: an update.Clin Chem Lab Med. 2008;46(11):1501-6. doi: 10.1515/CCLM.2008.291. Clin Chem Lab Med. 2008. PMID: 18778218 Review.
Cited by
-
Sensitive and quantitative detection of cardiac troponin I with upconverting nanoparticle lateral flow test with minimized interference.Sci Rep. 2021 Sep 21;11(1):18698. doi: 10.1038/s41598-021-98199-y. Sci Rep. 2021. PMID: 34548577 Free PMC article.
-
Evaluation of Molecularly Imprinted Polymers for Point-of-Care Testing for Cardiovascular Disease.Sensors (Basel). 2019 Aug 9;19(16):3485. doi: 10.3390/s19163485. Sensors (Basel). 2019. PMID: 31395843 Free PMC article. Review.
-
Measurement of cardiac troponins to detect myocardial infarction using high-sensitivity assays: South African guidelines.Cardiovasc J Afr. 2012 Nov;23(10):531-2. Cardiovasc J Afr. 2012. PMID: 23192256 Free PMC article. No abstract available.
-
Point-of-Care Compatibility of Ultra-Sensitive Detection Techniques for the Cardiac Biomarker Troponin I-Challenges and Potential Value.Biosensors (Basel). 2018 Nov 21;8(4):114. doi: 10.3390/bios8040114. Biosensors (Basel). 2018. PMID: 30469415 Free PMC article. Review.
-
High-sensitivity assays for troponin in patients with cardiac disease.Nat Rev Cardiol. 2017 Aug;14(8):472-483. doi: 10.1038/nrcardio.2017.48. Epub 2017 Apr 6. Nat Rev Cardiol. 2017. PMID: 28383022 Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Research Materials