Rotationally asymmetric multifocal IOL implantation with and without capsular tension ring: refractive and visual outcomes and intraocular optical performance
- PMID: 22496436
- DOI: 10.3928/1081597X-20120314-01
Rotationally asymmetric multifocal IOL implantation with and without capsular tension ring: refractive and visual outcomes and intraocular optical performance
Abstract
Purpose: To ascertain whether the refractive, visual, and intraocular optical quality outcomes of a rotationally asymmetric multifocal intraocular lens (IOL) are enhanced by the use of a capsular tension ring.
Methods: Ninety consecutive eyes from 53 patients (age range: 36 to 82 years) were divided into two groups: the no ring group comprised 43 eyes implanted with the multifocal rotationally asymmetric Lentis Mplus LS-312 (Oculentis GmbH) without a capsular tension ring; and the ring group comprised 47 eyes with the same IOL with a capsular tension ring. Distance and near visual acuity and refractive outcomes were evaluated pre- and postoperatively. Contrast sensitivity, intraocular aberrations, and defocus curve were evaluated postoperatively.
Results: Significant postoperative differences between groups were found in sphere, spherical equivalent refraction, and near addition (P<.02). Regarding defocus curve, significantly better visual acuity was present in eyes with the capsular tension ring for intermediate vision conditions (P<.05). Intraocular aberrometry did not differ significantly between groups (P<.09).
Conclusions: Refractive predictability and intermediate visual outcomes with the Lentis Mplus LS-312 IOL improved significantly when implanted in combination with a capsular tension ring.
Copyright 2012, SLACK Incorporated.