Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2012 May;9(5):468-72.
doi: 10.3171/2012.1.PEDS11110.

Experience with a gravitational valve in the management of symptomatic overdrainage in children with shunts

Affiliations

Experience with a gravitational valve in the management of symptomatic overdrainage in children with shunts

Martin R Weinzierl et al. J Neurosurg Pediatr. 2012 May.

Abstract

Object: Symptomatic overdrainage in children with shunt-treated hydrocephalus represents one of the more difficult shunt-related diseases and may require repeated surgery. Gravity-assisted valve design has become a standard device to avoid overdrainage in many European pediatric hydrocephalus centers. However, the use of a gravitational valve for relieving symptoms associated with overdrainage has not yet been addressed. The goal of this study was to evaluate the effectiveness of a gravitational valve in the treatment of symptomatic overdrainage in children with shunts.

Methods: Seventeen children with an adjustable shunt system and symptomatic overdrainage were treated by insertion of a gravitational valve. Clinical and radiological outcome were monitored for a minimum of 12 months after surgery.

Results: Implantation of a gravitational valve resulted in either resolution (n = 12) or improvement (n = 5) of the symptoms. In 1 patient, symptoms remained almost unchanged and the gravitational valve had to be upgraded, resulting in resolution of the symptoms. During follow-up, the pressure setting of the adjustable differential pressure valve had to be changed in 7 patients.

Conclusions: The gravitational valve was effective in improving symptomatic overdrainage in the majority of patients in the present study. Because the ideal pressure setting for a given patient is hard to determine a priori, adjustable valve systems appear to be beneficial.