Do pocket-sized ultrasound machines have the potential to be used as a tool to triage patients in obstetrics and gynecology?
- PMID: 22605511
- DOI: 10.1002/uog.11184
Do pocket-sized ultrasound machines have the potential to be used as a tool to triage patients in obstetrics and gynecology?
Abstract
Objectives: To evaluate the performance and potential impact on patient management of a pocket-sized ultrasound machine (PUM) in comparison to high-specification ultrasound machines (HSUM).
Methods: This was an observational cohort study with 204 unselected patients in three categories: 1) women with pain and bleeding in early pregnancy (101 patients); 2) women presenting for routine obstetric ultrasound assessment (53 patients); 3) women with possible gynecological pathology (50 patients). Scans were carried out transabdominally using a PUM. A second operator repeated the examination transvaginally and/or transabdominally, depending on the clinical indication, using an HSUM. The operators were blind to each other's findings.
Results: In the early pregnancy group, there was good to very good agreement between the PUM and HSUM for identifying the presence or absence of an embryo, gestational sac, fetal heart motion, pregnancy location and final diagnostic outcome (kappa coefficients, 0.844, 0.843, 0.729, 0.785 and 0.812, respectively; P < 0.0001). In the obstetric ultrasound group there was good to very good agreement for fetal presentation, placental location and placental position (kappa coefficients, 0.924, 0.924 and 0.647, respectively; P < 0.0001). In the gynecological pathology group, there was very good agreement for final diagnosis and type of ovarian mass (low risk or complex) (kappa coefficients, 0.846 and 0.930, respectively; P < 0.0001). For the measured continuous variables, there was close agreement for crown-rump length, mean sac diameter, femur length and mean diameter of an ovarian mass, but not for endometrial thickness. Neither patient demographics (age, body mass index, ethnicity) nor operator experience and familiarity with a PUM had an impact on agreement between the two machines. If a PUM had been the only equipment available for an initial assessment, only two cases would have led to a suboptimal patient management plan.
Conclusion: The findings and final diagnosis in the three study groups were similar for both a PUM used transabdominally and an HSUM used transvaginally and/or transabdominally.
Copyright © 2012 ISUOG. Published by John Wiley & Sons, Ltd.
Similar articles
-
A futuristic vision of pocket ultrasound machines: watch this space.Australas J Ultrasound Med. 2014 Aug;17(3):110-112. doi: 10.1002/j.2205-0140.2014.tb00025.x. Epub 2015 Dec 31. Australas J Ultrasound Med. 2014. PMID: 28191219 Free PMC article. Review.
-
Use of a pocket-sized ultrasound machine (PUM) for routine examinations in the third trimester of pregnancy.Ultrasound Obstet Gynecol. 2014 Jul;44(1):64-8. doi: 10.1002/uog.13285. Ultrasound Obstet Gynecol. 2014. PMID: 24357339
-
Accuracy of portable ultrasound machines for obstetric biometry.Ultrasound Obstet Gynecol. 2024 Jun;63(6):772-780. doi: 10.1002/uog.27541. Ultrasound Obstet Gynecol. 2024. PMID: 38011589 Free PMC article.
-
Assessment of fetal anatomy at the 11-14-week ultrasound examination.Ultrasound Obstet Gynecol. 2004 Dec;24(7):730-4. doi: 10.1002/uog.1775. Ultrasound Obstet Gynecol. 2004. PMID: 15586371
-
Practical and financial considerations that affect selection and purchase of ultrasound equipment.Obstet Gynecol Clin North Am. 1998 Sep;25(3):663-75. doi: 10.1016/s0889-8545(05)70033-4. Obstet Gynecol Clin North Am. 1998. PMID: 9710917 Review.
Cited by
-
Color Doppler Ultrasonography to Evaluate Hypoechoic Areas in Pressure Ulcers: A Report of Two Cases.J Med Ultrasound. 2018 Jul-Sep;26(3):163-165. doi: 10.4103/JMU.JMU_60_18. Epub 2018 Sep 14. J Med Ultrasound. 2018. PMID: 30283205 Free PMC article.
-
Hand-Held Ultrasound Devices Compared with High-End Ultrasound Systems: A Systematic Review.Diagnostics (Basel). 2019 Jun 15;9(2):61. doi: 10.3390/diagnostics9020061. Diagnostics (Basel). 2019. PMID: 31208078 Free PMC article. Review.
-
Pocket ultrasound device as a complement to physical examination for ascites evaluation and guided paracentesis.Intern Emerg Med. 2016 Apr;11(3):461-6. doi: 10.1007/s11739-016-1406-x. Epub 2016 Feb 19. Intern Emerg Med. 2016. PMID: 26895032
-
A futuristic vision of pocket ultrasound machines: watch this space.Australas J Ultrasound Med. 2014 Aug;17(3):110-112. doi: 10.1002/j.2205-0140.2014.tb00025.x. Epub 2015 Dec 31. Australas J Ultrasound Med. 2014. PMID: 28191219 Free PMC article. Review.
-
Diagnoses of postpartum urinary retention using next-generation non-piezo ultrasound technology: assessing the accuracy and benefits.Sci Rep. 2024 Dec 30;14(1):31844. doi: 10.1038/s41598-024-83160-6. Sci Rep. 2024. PMID: 39738356 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Research Materials
Miscellaneous