Successful nanolitre real-time PCR detection of respiratory pathogens in clinical specimens
- PMID: 22630162
- DOI: 10.1111/j.1469-0691.2012.03896.x
Successful nanolitre real-time PCR detection of respiratory pathogens in clinical specimens
Abstract
We performed a proof-of-concept study to determine if human pathogens could be detected in clinical specimens using nanolitre-volume real-time PCR. Nanolitre PCR for Bordetella pertussis/B. parapertussis and respiratory syncytial virus (RSV) was performed on nasopharyngeal specimens and results compared with conventional methods. B. pertussis/B. parapertussis nanolitre PCR detection was 100% sensitive (20/20; 95% CI, 84-100%) and 100% specific (26/26; 95% CI, 87-100%). RSV nanolitre PCR was also 100% sensitive (21/21; 95% CI, 85-100%) and specific (25/25; 95%, CI 87-100%). Respiratory pathogens can be successfully detected in clinical specimens using nanolitre-volume PCR.
© 2012 The Authors. Clinical Microbiology and Infection © 2012 European Society of Clinical Microbiology and Infectious Diseases.