ST waveform analysis versus cardiotocography alone for intrapartum fetal monitoring: a meta-analysis of randomized trials
- PMID: 22639349
- DOI: 10.1055/s-0032-1314886
ST waveform analysis versus cardiotocography alone for intrapartum fetal monitoring: a meta-analysis of randomized trials
Abstract
Objective: To estimate the effectiveness of intrapartum ST waveform analysis (STAN) versus cardiotocography (CTG) alone in prevention of metabolic acidosis.
Study design: Meta-analysis of randomized trials comparing intrapartum fetal monitoring utilizing STAN with CTG versus CTG alone. Primary outcome was neonatal metabolic acidosis, defined as umbilical arterial pH <7.05 and base deficit >12 mmol/L.
Results: Five randomized trials including 15,303 singletons, vertex, term, or near-term pregnancies met inclusion criteria and were analyzed. Compared with CTG alone, STAN with CTG was associated with similar incidences of metabolic acidosis (0.81% versus 1.12%, relative risk [RR] 0.80; 95% confidence interval [CI] 0.44 to 1.47), perinatal death, neonatal encephalopathy, Apgar score <7 at 5 minutes, admission to neonatal intensive care unit, and cesarean delivery. Operative vaginal delivery (OVD) was lower in the STAN with CTG compared with CTG alone (13.56% versus 15.20%; RR 0.89; 95% CI 0.83 to 0.97).
Conclusion: There is no difference in perinatal outcomes between STAN with CTG compared with CTG alone, except for lower rate of OVD.
Thieme Medical Publishers 333 Seventh Avenue, New York, NY 10001, USA.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous