Effect of processing methods for transmission electron microscopy on corneal collagen fibrils diameter and spacing

Affiliations

Effect of processing methods for transmission electron microscopy on corneal collagen fibrils diameter and spacing

Saeed Akhtar. Microsc Res Tech. 2012 Oct.

Abstract

Introduction: The corneal tissue was processed in fixatives and embedded in resin for transmission electron microscopy to observe the ultrastructure of the collagen fibrils (CFs). The effect of these processing methods on the CF diameter and the interfibrillar spacing was studied.

Methods: Four normal human corneal buttons were used for this study. A part of each cornea was fixed in 2.5% glutaraldehyde containing cuprolinic blue in sodium acetate buffer and embedded in spurr's resin (SpurrCB). A second part of each cornea was fixed in 2.5% glutaraldehyde + osmium tetroxide and embedded spurr's resin (SpurrOsm). The third part of each cornea was fixed in paraformaldehyde (4%) and embedded in LR White at 4°C (LRWhite). Ultrathin sections were stained with uranyl acetate and lead citrate.

Results: In the tissue, fixed in SpurrCB, the diameter was 38.4 ± 5.9 nm and spacing between CF was 52.5 ± 5.3 nm. In the tissue fixed in SpurrOsm, the diameter was 28.37 ± 5.84 nm and spacing between CF was 45 ± 4.57 nm. In the tissue fixed in LR White, the CF diameter was 24 ± 2.3 nm and spacing between CF was 39.0 ± 4.2 nm. The diameters and interfibrillar spacing of the tissue processed by SpurrCB, SpurrOsm, and LRWhite were significantly different (P < 0.001) from one another.

Conclusion: Our study shows that there is a variation in the CF diameter and spacing depending on the method of fixation and embedding resins used. This needs to be considered when comparative studies using different methods are done.