Clinical Trial
. 2012 May-Jun;64(3):273-9.
doi: 10.1016/S0019-4832(12)60086-8.

Systemic exposure of sirolimus after coronary stent implantation in patients with de novo coronary lesions: Supralimus-Core® pharmacokinetic study

Affiliations
Clinical Trial

Systemic exposure of sirolimus after coronary stent implantation in patients with de novo coronary lesions: Supralimus-Core® pharmacokinetic study

Ashok S Thakkar et al. Indian Heart J. 2012 May-Jun.

Abstract

Objectives: This study was conducted to assess the systemic drug release and distribution of sirolimus-eluting coronary stents.

Methods: Twenty patients with coronary artery disease (CAD) were treated with 1, 2, or 3 a newly designed metallic stents. Blood samples were drawn at 14 time points to determine the pharmacokinetic of sirolimus. Whole blood concentrations of sirolimus were determined by using a sensitive validated high-performance liquid chromatography mass spectrometry/mass spectrometry method.

Results: Minimal measurable blood levels were detectable at 7 days. Across all dose levels, individual T(max) values ranged from 1.00 hour and 12.00 hours; individual C(max) ranged from 0.73 ng/mL and 4.13 ng/mL.

Conclusion: This study confirms the limited exposure of the systemic circulation of the eluted drug with the use of the Supralimus-Core® Sirolimus-Eluting Coronary Stent System (Sahajanand Medical Technologies Pvt. Ltd., Surat, India). In this study, sirolimus concentration in systemic circulation is to be safe, well-tolerated and short-lived.

PubMed Disclaimer

References

    1. Bertrand OF, Sipehia R, Mongrain R. Biocompatibility aspects of new stent technology. J Am Coll Cardiol. 1998;32:562–571. - PubMed
    1. Hwang CW, Wu D, Edelman ER. Physiological transport forces govern drug distribution for stent-based delivery. Circulation. 2001;104:600–605. - PubMed
    1. Creel CJ, Lovich MA, Edelman ER. Arterial paclitaxel distribution and deposition. Circ Res. 2000;86:879–884. - PubMed
    1. Fajadet J, Morice MC, Bode C. Maintenance of long-term clinical benefit with sirolimus-eluting coronary stents: three-year results of the RAVEL trial. Circulation. 2005;111:1040–1044. - PubMed
    1. Holmes DR, Jr, Leon MB, Moses JW. Analysis of 1-year clinical outcomes in the SIRIUS trial: a randomized trial of a sirolimus-eluting stent versus a standard stent in patients at high risk for coronary restenosis. Circulation. 2004;109:634–640. - PubMed

Publication types

LinkOut - more resources