Interrater and intrarater reliability of photoplethysmography for measuring toe blood pressure and toe-brachial index in people with diabetes mellitus

Affiliations

Interrater and intrarater reliability of photoplethysmography for measuring toe blood pressure and toe-brachial index in people with diabetes mellitus

Christopher Scanlon et al. J Foot Ankle Res. .

Abstract

Background: A reliable tool to measure arterial flow to the feet in people with diabetes is required given that they are particularly prone to peripheral arterial disease. Traditionally, the ankle brachial index (ABI) has been used to measure arterial circulation, but its application is limited due to calcification of larger arteries. More recently, toe pressure and the toe brachial index (TBI) has been suggested as superior to ABI measurements because they assess smaller digital arteries less prone to arterial calcification. However, reliability studies for the clinical use of photoplethysmography (PPG) in people with diabetes are lacking.

Methods: Sixty people with diabetes mellitus (35 males and 25 females, mean age 59.6 yrs) consented to take part in the study. The majority (92%) had type 2 diabetes and 8% had type 1 diabetes. Forty-three percent were diagnosed as having peripheral neuropathy when tested using a biothesiometer and 15% were current smokers (10 - 40/day). A podiatrist and a diabetes educator measured toe and brachial blood pressure independently and in a random order using PPG. These measurements were repeated again seven days later, and subsequently analysed with intraclass correlation coefficients (ICC), 95% confidence intervals (CI) and standard error of measurement (SEM).

Results: The intrarater reliability of measuring toe pressures was excellent (ICC3,1 =0.78-0.79, SEM 8 mmHg) and interrater reliability was also excellent (ICC2,2 = 0.93, SEM 4 mmHg). The intrarater reliability for measuring brachial pressures was generally poor (ICC3,1 = 0.40 - 0.42, SEM 19 mmHg) and interrater reliability was fair-good (ICC2,2. 0.65, SEM 14 mmHg). The TBI intrarater reliability was fair-good (ICC3,1 = 0.51-0.72, SEM 0.08), whilst the interrater reliability of TBI was excellent (ICC2,2 = 0.85, SEM 0.07).

Conclusions: Based on these results, interrater and intrarater reliability of PPG is excellent for measuring toe blood pressure, good for TBI and only fair for brachial pressures in people with diabetes mellitus.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Khan NA, Rahim SA, Anand SS, Simel DL, Panju A. Does the clinical examination predict lower extremity peripheral arterial disease? JAMA. 2006;295:536–546. doi: 10.1001/jama.295.5.536. - DOI - PubMed
    1. de Graaff JC, Ubbink DT, Legemate DA, de Haan RJ, Jacobs MJ. Interobserver and intraobserver reproducibility of peripheral blood and oxygen pressure measurements in the assessment of lower extremity arterial disease. J Vasc Surg. 2001;33:1033–1040. doi: 10.1067/mva.2001.108011. - DOI - PubMed
    1. Ramanathan A, Conaghan PJ, Jenkinson AD, Bishop CR. Comparison of ankle-brachial pressure index measurements using an automated oscillometric device with the standard Doppler ultrasound technique. ANZ J Surg. 2003;73:105–108. doi: 10.1046/j.1445-2197.2003.02582.x. - DOI - PubMed
    1. Yamada T, Ohta T, Ishibashi H, Sugimoto I, Iwata H, Takahashi M, Kawanishi J. Clinical reliability and utility of skin perfusion pressure measurement in ischemic limbs–comparison with other noninvasive diagnostic methods. J Vasc Surg. 2008;47:318–323. doi: 10.1016/j.jvs.2007.10.045. - DOI - PubMed
    1. Weatherley BD, Chambless LE, Heiss G, Catellier DJ, Ellison CR. The reliability of the ankle-brachial index in the Atherosclerosis Risk in Communities (ARIC) study and the NHLBI Family Heart Study (FHS) BMC Cardiovasc Disord. 2006;6:7.. doi: 10.1186/1471-2261-6-7. - DOI - PMC - PubMed