Performance of atrial tachyarrhythmia-sensing algorithms in dual-chamber pacing using a fixed long AV delay in patients with sinus node dysfunction

Affiliations

Performance of atrial tachyarrhythmia-sensing algorithms in dual-chamber pacing using a fixed long AV delay in patients with sinus node dysfunction

Sami Pakarinen et al. J Interv Card Electrophysiol. 2012 Nov.

Abstract

Aims: The aim of the study was to evaluate the performance of pacemakers (PM) atrial tachyarrhythmia (AT)-sensing algorithms in sinus node dysfunction (SND) patients with DDDR pacing programmed with a fixed long atrioventricular (AV) delay.

Methods: In a prospective study, a total of 60 patients with SND were implanted with a dual-chamber PM with two different algorithms for detection of ATs. The study was done with a 3 month data collection period retrieved from the memory of PM and with a 7 day external Holter recording period.

Results: In 13 of 16 (81 %) patients whose Holter recording revealed the presence of ATs, episodes of AT sensing were retrieved from the PM memory with electrograms verifications, confirming that the devices had detected the ATs. Very short ATs seen in Holter recordings were missed by the PM with three patients. However, with all these patients after 3 months of follow up period, there were recognized periods of ATs by the PM. With ten (17 %) patients, there were intermittent periods of undersensing by the PM although continuous atrial fibrillation (AF) was seen in the Holter recording. Retrograde conduction caused false AT detection due to repetitive non-reentrant ventriculoatrial synchronous rhythm (RNRVAS) in six (25 %) of the 24 patients with retrograde conduction.

Conclusions: Even with long AV delay, ATs can be accurately identified. However, transient undersensing of continuous AF and non-detection of very short AT episodes can still occur. Programming a long AV delay predisposes to RNRVAS which can cause false AT detection and symptoms in SND patients who have retrograde conduction.

PubMed Disclaimer

References

    1. Pacing Clin Electrophysiol. 2004 Jun;27(6 Pt 1):824-6 - PubMed
    1. Clin Res Cardiol. 2009 Jan;98(1):25-32 - PubMed
    1. Pacing Clin Electrophysiol. 2005 Jul;28(7):620-9 - PubMed
    1. J Interv Card Electrophysiol. 2010 Dec;29(3):191-8 - PubMed
    1. J Cardiovasc Electrophysiol. 2006 Sep;17(9):992-7 - PubMed

Publication types

LinkOut - more resources