Comparative Study
. 2012 Aug;50(8):2732-6.
doi: 10.1128/JCM.05501-11. Epub 2012 Jun 12.

First evaluation of automated specimen inoculation for wound swab samples by use of the Previ Isola system compared to manual inoculation in a routine laboratory: finding a cost-effective and accurate approach

Affiliations
Comparative Study

First evaluation of automated specimen inoculation for wound swab samples by use of the Previ Isola system compared to manual inoculation in a routine laboratory: finding a cost-effective and accurate approach

Alexander Mischnik et al. J Clin Microbiol. 2012 Aug.

Abstract

Automation of plate streaking is ongoing in clinical microbiological laboratories, but evaluation for routine use is mostly open. In the present study, the recovery of microorganisms from the Previ Isola system plated polyurethane (PU) swab samples is compared to manually plated control viscose swab samples from wounds according to the CLSI procedure M40-A (quality control of microbiological transport systems). One hundred twelve paired samples (224 swabs) were analyzed. In 80/112 samples (71%), concordant culture results were obtained with the two methods. In 32/112 samples (29%), CFU recovery of microorganisms from the two methods was discordant. In 24 (75%) of the 32 paired samples with a discordant result, Previ Isola plated PU swabs were superior. In 8 (25%) of the 32 paired samples with a discordant result, control viscose swabs were superior. The quality of colony growth on culture media for further investigations was superior with Previ Isola inoculated plates compared to manual plating techniques. Gram stain results were concordant between the two methods in 62/112 samples (55%). In 50/112 samples (45%), the results of Gram staining were discordant between the two methods. In 34 (68%) of the 50 paired samples with discordant results, Gram staining of PU swabs was superior to that of control viscose swabs. In 16 (32%) of the 50 paired samples, Gram staining of control viscose swabs was superior to that of PU swabs. We report the first clinical evaluation of Previ Isola automated specimen inoculation for wound swab samples. This study suggests that use of an automated specimen inoculation system has good results with regard to CFU recovery, quality of Gram staining, and accuracy of diagnosis.

PubMed Disclaimer

Figures

Fig 1
Fig 1
Colony growth on Columbia agar, chocolate agar, and MacConkey agar (top to bottom, respectively) streaked manually (left) and corresponding agar plates streaked by Previ Isola (right).

References

    1. Bourbeau PP, Swartz BL. 2009. First evaluation of the WASP, a new automated microbiology plating instrument. J. Clin. Microbiol. 47:1101–1106 - PMC - PubMed
    1. Brook I. 1987. Comparison of two transport systems for recovery of aerobic and anaerobic bacteria from abscesses. J. Clin. Microbiol. 25:2020–2022 - PMC - PubMed
    1. Bruno LC, Janda WM, Villarreal RI, Nguyen AK. 2011. Evaluation of bioMerieux PREVI Isola automated plate streaker, poster 2152. Abstr. 111th Gen. Meet. Am. Soc. Microbiol. http://gm.asm.org/
    1. Citron DM, Warren YA, Hudspeth MK, Goldstein EJ. 2000. Survival of aerobic and anaerobic bacteria in purulent clinical specimens maintained in the Copan Venturi Transystem and Becton Dickinson Port-a-Cul transport systems. J. Clin. Microbiol. 38:892–894 - PMC - PubMed
    1. Collee JG. 1980. Factors contributing to loss of anaerobic bacteria in transit from the patient to the laboratory. Infection 8(Suppl. 2):S145–S147 - PubMed

MeSH terms