Intraoperative characterization of interventricular mechanical dyssynchrony using electroanatomic mapping system--a feasibility study
- PMID: 22695763
- DOI: 10.1007/s10840-012-9695-0
Intraoperative characterization of interventricular mechanical dyssynchrony using electroanatomic mapping system--a feasibility study
Abstract
Background: Interventricular mechanical dyssynchrony (VVMD) is a strong predictor of cardiac resynchronization therapy (CRT) response. However, no simple and reliable clinical method of measuring VVMD during CRT implant is currently available. We tested the hypothesis that the EnSite™ NavX™ system (St. Jude Medical, St. Paul, MN, USA) can be used intraoperatively to determine VVMD, thereby facilitating CRT optimization.
Methods: During CRT implant, the leads in the right atrium (RA), right ventricle (RV), and left ventricle (LV) were connected to the EnSite™ NavX™ system to record the real-time 3D motion of the lead electrodes. The distances from RA to RV lead electrodes (RA-RV) and RA to LV lead electrodes (RA-LV) were computed over ten cardiac cycles during each of RV pacing and biventricular (BiV) pacing, respectively. The degree of synchrony was computed from the distance waveforms between RA-RV and RA-LV by a cross-covariance method to characterize VVMD. Septal-to-posterior wall motion delay (SPWMD) from M-mode echocardiography (echo) was measured for reference at each pacing intervention. VVMD was present in all five patients undergoing CRT implant.
Results: Four of the five patients demonstrated clear improvement in EnSite™ NavX™-derived VVMD during BiV versus RV pacing, which corresponded to the SPWMD results by echo.
Conclusions: It is feasible to characterize VVMD and resynchronization in CRT patients with the EnSite™ NavX™ system during implant, demonstrating its potential as a tool for intraoperative CRT optimization.
Trial registration: ClinicalTrials.gov NCT00632450.
Similar articles
-
Three-Dimensional Cardiac Mapping Characterizes Ventricular Contractile Patterns during Cardiac Resynchronization Therapy Implant: A Feasibility Study.Pacing Clin Electrophysiol. 2015 Sep;38(9):1091-8. doi: 10.1111/pace.12674. Epub 2015 Jul 16. Pacing Clin Electrophysiol. 2015. PMID: 26096125 Clinical Trial.
-
CRT in a congenital heart disease patient with interventricular dyssynchrony due to an RV conduction delay.Pacing Clin Electrophysiol. 2020 Sep;43(9):1051-1057. doi: 10.1111/pace.13978. Epub 2020 Jul 2. Pacing Clin Electrophysiol. 2020. PMID: 32519349
-
Three-dimensional electroanatomic mapping of the coronary veins during cardiac resynchronization therapy implant: feasibility and possible applications.J Interv Card Electrophysiol. 2014 Nov;41(2):147-53. doi: 10.1007/s10840-014-9932-9. Epub 2014 Jul 9. J Interv Card Electrophysiol. 2014. PMID: 25005455
-
Usefulness of the 12-lead electrocardiogram in the follow-up of patients with cardiac resynchronization devices. Part II.Cardiol J. 2011;18(6):610-24. doi: 10.5603/cj.2011.0024. Cardiol J. 2011. PMID: 22113748 Review.
-
Assessment of mechanical dyssynchrony in cardiac resynchronization therapy.Dan Med J. 2014 Dec;61(12):B4981. Dan Med J. 2014. PMID: 25441737 Review.