Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Clinical Trial
. 2013 Oct;21(10):2194-9.
doi: 10.1007/s00167-012-2098-z. Epub 2012 Jun 27.

Evaluation of the accuracy of a patient-specific instrumentation by navigation

Affiliations
Clinical Trial

Evaluation of the accuracy of a patient-specific instrumentation by navigation

Fabio Conteduca et al. Knee Surg Sports Traumatol Arthrosc. 2013 Oct.

Abstract

Purpose: The aim of this study is to evaluate the accuracy of a patient-specific instrumentation (PSI) as assessed by the intraoperative use of knee navigation software during the surgical procedure.

Methods: Fifteen patients with primary gonarthrosis were selected for unilateral total knee arthroplasty. The first three patients were excluded from this study, as they were considered to be a warm up to set-up the procedure. All patients were operated on with a cemented posterior-stabilised prosthesis cruciate ligament-sacrificing by the same surgeon using the patient matched cutting jigs. The size of the implant, level of resection, and alignment in the coronal and sagittal planes were evaluated. An unsatisfactory result was considered an error ≥2° in both planes for each component as a possible error of 4° could result in aggravation.

Results: On the coronal plane the mean deviation of the tibial guide from the ideal alignment was 1.2 ± 1.5 (range 0-5°) and in the sagittal plane was 3.8 ± 2.4 (range 0-7.5°). On the coronal plane the mean deviation of the femoral guide from the ideal alignment was 1.2 ± 0.6 and in the sagittal was 3.7 ± 2.

Conclusion: On the basis of this preliminary experience the PSI system based only on data acquisition with A-P radiograms and RMN cannot be defined as accurate. In cases of the use of the custom made cutting jigs it is recommended to perform an accurate control of the alignment before making the cuts, for any step of the procedure.

Level of evidence: II.

PubMed Disclaimer

References

    1. J Arthroplasty. 2006 Jun;21(4 Suppl 1):29-32 - PubMed
    1. Knee. 2012 Aug;19(4):237-45 - PubMed
    1. Clin Orthop Relat Res. 1998 Nov;(356):144-53 - PubMed
    1. Knee Surg Sports Traumatol Arthrosc. 2012 Jul;20(7):1307-22 - PubMed
    1. Clin Orthop Relat Res. 2006 Mar;444:184-92 - PubMed

Publication types

MeSH terms

LinkOut - more resources