Validation of a new protocol for ¹⁸F-FDG infusion using an automatic combined dispenser and injector system
- PMID: 22740098
- DOI: 10.1007/s00259-012-2174-0
Validation of a new protocol for ¹⁸F-FDG infusion using an automatic combined dispenser and injector system
Abstract
Purpose: In nuclear medicine, radiopharmaceuticals are usually administered in unit doses partitioned from multi-dose vials. The partitioning typically takes place in a radiopharmacy, depending on local practice. Automatic, as opposed to manual, partitioning and administration should reduce radiation exposure of the personnel involved, improve the accuracy of the administration and mitigate contamination. This study set out to verify and validate the (18)F-fluorodeoxyglucose (FDG) administration procedure performed using Intego (MEDRAD, Inc., Warrendale, PA, USA), a combined dispenser and injector system. We considered maintenance of sterility and the system's potential to improve, with respect to the manual procedure, the accuracy of net administered (18)F-FDG radioactivity in patients and the radiation protection of operators.
Methods: A media-fill procedure was used to assess whether sterility is maintained during use of the Intego system. Simulating a typical working day's setup and use of the system, we investigated the accuracy of the net administered (18)F-FDG activity obtained with Intego versus the manual dose delivery system. We also measured personnel radiation exposure during use of Intego and during manual administration and recorded and compared environmental doses in the two conditions.
Results: The radiopharmaceutical remained sterile in all the tests performed. The accuracy of the net (18)F-FDG activity delivered to the patients was found to be within 3 % points, as required by European Association of Nuclear Medicine (EANM) guidelines on (18)F-FDG imaging procedures. With Intego, the residual radioactivity in the tubing was 0.20 MBq, corresponding to approximately 0.07 % of the mean activity delivered. With manual injection, the residual radioactivity in the syringe averaged 7.37 MBq, corresponding to a mean error of 2.9 % in the delivered dose. During the injection step of the positron emission tomography (PET) procedure, whole-body and extremity radiation exposures were significantly reduced with Intego by 38 and by 94 %, respectively, compared to the levels associated with manual administration (p < 0.05).
Conclusion: Integoaccurately partitions and administers sterile doses of (18)F-FDG from multi-dose vials. Compared with standard manual (18)F-FDG administration, the new procedure with an automatic dispensing and injection system greatly reduces the extremity dose to the operator involved in the administration of the radiopharmaceutical.
Similar articles
-
Occupational radiation dosimetry assessment using an automated infusion device for positron-emitting radiotracers.J Nucl Med Technol. 2012 Dec;40(4):244-8. doi: 10.2967/jnmt.112.106070. Epub 2012 Sep 27. J Nucl Med Technol. 2012. PMID: 23024340
-
Evaluation of an automated FDG dose infuser to PET-CT patients.Radiat Prot Dosimetry. 2015 Jul;165(1-4):457-60. doi: 10.1093/rpd/ncv077. Epub 2015 Mar 26. Radiat Prot Dosimetry. 2015. PMID: 25821215
-
Validation of an automated dose-dispensing system for 18F-FDG administrations and associated reduction in operator extremity dose.Nucl Med Commun. 2014 Feb;35(2):151-9. doi: 10.1097/MNM.0000000000000027. Nucl Med Commun. 2014. PMID: 24217428
-
The role of ¹⁸F-FDG PET or ¹⁸F-FDG-PET/CT in the evaluation of solitary pulmonary nodules.Eur J Radiol. 2015 Oct;84(10):2032-7. doi: 10.1016/j.ejrad.2015.06.008. Epub 2015 Jun 11. Eur J Radiol. 2015. PMID: 26094866 Review.
-
Delivery Methods of Radiopharmaceuticals: Exploring Global Strategies to Minimize Occupational Radiation Exposure.J Nucl Med Technol. 2025 Mar 5;53(1):2-6. doi: 10.2967/jnmt.124.268043. J Nucl Med Technol. 2025. PMID: 39532489 Review.
Cited by
-
Quality Assessment in FDG-PET/CT Imaging of Head-and-Neck Cancer: One Home Run Is Better Than Two Doubles.Front Oncol. 2020 Aug 14;10:1458. doi: 10.3389/fonc.2020.01458. eCollection 2020. Front Oncol. 2020. PMID: 32923399 Free PMC article. Review.
-
An account of data entry inconsistencies and their impact on positron emission tomography quantification.Medicine (Baltimore). 2018 Sep;97(37):e12312. doi: 10.1097/MD.0000000000012312. Medicine (Baltimore). 2018. PMID: 30212971 Free PMC article.
-
Clinical validation of a population-based input function for 20-min dynamic whole-body 18F-FDG multiparametric PET imaging.EJNMMI Phys. 2022 Sep 8;9(1):60. doi: 10.1186/s40658-022-00490-y. EJNMMI Phys. 2022. PMID: 36076097 Free PMC article.
-
Effective and equivalent dose minimization for personnel in PET procedures: how far are we from the goal?Eur J Nucl Med Mol Imaging. 2016 Dec;43(13):2279-2282. doi: 10.1007/s00259-016-3513-3. Eur J Nucl Med Mol Imaging. 2016. PMID: 27624191 No abstract available.
-
ARAS: an automated radioactivity aliquoting system for dispensing solutions containing positron-emitting radioisotopes.EJNMMI Res. 2016 Dec;6(1):22. doi: 10.1186/s13550-016-0176-9. Epub 2016 Mar 9. EJNMMI Res. 2016. PMID: 26956805 Free PMC article.
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources