Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Randomized Controlled Trial
. 2012;7(6):e38277.
doi: 10.1371/journal.pone.0038277. Epub 2012 Jun 22.

Use of an atrial lead with very short tip-to-ring spacing avoids oversensing of far-field R-wave

Collaborators, Affiliations
Randomized Controlled Trial

Use of an atrial lead with very short tip-to-ring spacing avoids oversensing of far-field R-wave

Christof Kolb et al. PLoS One. 2012.

Abstract

Objective: The AVOID-FFS (Avoidance of Far-Field R-wave Sensing) study aimed to investigate whether an atrial lead with a very short tip-to-ring spacing without optimization of pacemaker settings shows equally low incidence of far-field R-wave sensing (FFS) when compared to a conventional atrial lead in combination with optimization of the programming.

Methods: Patients receiving a dual chamber pacemaker were randomly assigned to receive an atrial lead with a tip-to-ring spacing of 1.1 mm or a lead with a conventional tip-to-ring spacing of 10 mm. Postventricular atrial blanking (PVAB) was programmed to the shortest possible value of 60 ms in the study group, and to an individually determined optimized value in the control group. Atrial sensing threshold was programmed to 0.3 mV in both groups. False positive mode switch caused by FFS was evaluated at one and three months post implantation.

Results: A total of 204 patients (121 male; age 73±10 years) were included in the study. False positive mode switch caused by FFS was detected in one (1%) patient of the study group and two (2%) patients of the control group (p = 0.62).

Conclusion: The use of an atrial electrode with a very short tip-to-ring spacing avoids inappropriate mode switch caused by FFS without the need for individual PVAB optimization.

Trial registration: ClinicalTrials.gov NCT00512915.

PubMed Disclaimer

Conflict of interest statement

Competing Interests: This study was financially supported by St. Jude Medical, Eschborn, Germany. CK has received lecture fees from Biotronik, Boston, Medtronic, Sorin, and St. Jude Medical, and is or was an advisor to Biotronik, Sorin and St. Jude Medical. GN and KG received lecture fees from Biotronik, Medtronic, Sorin and St. Jude Medical. GN received support for travelling when presenting study data from St. Jude Medical. UL has received lecture fees from Medtronic and St. Jude Medical. All other authors have declared that no competing interests exist. This does not alter the authors’ adherence to all the PLoS ONE policies on sharing data and materials.

Figures

Figure 1
Figure 1. Study flow-chart.
Figure 2
Figure 2. Inappropriate mode switch due to FFS: Stored episode depicting inappropriate mode switch due to FFS from the control group.
Despite optimized PVAB of 140 ms inappropriate mode switch occurs because the coupling interval of the far-field R-wave is 160 ms. First line: bipolar atrial electrogram, shows bipolar atrial, second line: bipolar ventricular electrogram, bottom line: marker channel with AMS = mode switch; AP = atrial pacing; AS = atrial sensed event; AS on black background = atrial sensed event in refractory period; VP = ventricular pacing.
Figure 3
Figure 3. Distribution of programmed PVAB during follow-up by randomization groups.
Nine patients of the study group initially received a PVAB that was erroneously programmed to a value other than 60 ms which was corrected after one month in eight of the patients and remained prolonged in one patient. None of these patients had inappropriate mode switch due to FFS. One patient of the study group patient exhibited inappropriate mode switch due to FFS and the PVAB was then individually optimized (150 ms). For the control group PVAB is shown as determined to be optimal at discharge. In case of changes of the PVAB programming during the follow-up, the longest programmed PVAB for both groups are displayed.

References

    1. Fitts SM, Hill MR, Mehra R, Gillis AM. High rate atrial tachyarrhythmia detections in implantable pulse generators: low incidence of false-positive detections. The PA Clinical Trial Investigators. Pacing Clin Electrophysiol. 2000;23:1080–6. - PubMed
    1. Pürerfellner H, Gillis AM, Holbrook R, Hettrick DA. Accuracy of atrial tachyarrhythmia detection in implantable devices with arrhythmia therapies. Pacing Clin Electrophysiol. 2004;27:983–92. - PubMed
    1. De Voogt WG, van Hemel NM, van de Bos AA, Koistinen J, Fast JH. Verification of pacemaker automatic mode switching for the detection of atrial fibrillation and atrial tachycardia with Holter recording. Europace. 2006;8:950–61. - PubMed
    1. De Voogt WG, van Hemel NM. Clinical application of pacemakers in atrial tachyarrhythmias. Neth Heart J. 2008. pp. S20–4. - PMC - PubMed
    1. Glotzer TV, Hellkamp AS, Zimmerman J, Sweeney MO, Yee R, et al. Atrial high rate episodes detected by pacemaker diagnostics predict death and stroke. Circulation. 2003;107:1614–9. - PubMed

Publication types

Associated data