Torsion of monofilament and polyfilament sutures under tension decreases suture strength and increases risk of suture fracture
- PMID: 22762905
- DOI: 10.1016/j.jmbbm.2012.02.001
Torsion of monofilament and polyfilament sutures under tension decreases suture strength and increases risk of suture fracture
Abstract
Background: A continuous running suture is the preferential method for abdominal closure. In this technique the suture is secured with an initial knot and successive tissue bites are taken. At each tissue bite, the needle is rotated through the tissue; in doing so, the suture can twist around the knot which acts as an anchor.
Objective: To determine the effect of axial torsional forces on sutures used in abdominal closure.
Methods: The effect of axial twisting on polydioxanone (PDS*II), polyglactin (Vicryl), polypropylene (Prolene) and nylon (Ethilon) sutures was investigated using a uniaxial testing device.
Results: The maximum tensile force withstood for untwisted sutures was determined: polydioxanone failed at a tensile force of 116.4±0.84 N, polyglactin failed at 113.9±2.4 N, polypropylene failed at 71.1±1.5 N and nylon failed at 61.8±0.5 N. Twisting decreased the maximum tensile force of all sutures; one complete twist per 10 mm (i.e., 15 twists) decreased the tensile strength of polydioxanone by 21%, polyglactin by 23%, polypropylene by 16% and nylon by 13%, p<0.001. Excessive twisting caused a nonlinear decrease in suture strength, with one twist per 75 mm (i.e., 20 twists) of polydioxanone decreasing strength by 39%, P<0.001.
Conclusion: The effect of excessive twisting on the mechanical properties of sutures is a previously unrecognised phenomenon. Surgeons should be aware that this can result in a decrease in suture strength and reduce the elasticity of the material, and therefore need to adapt their practice to reduce the torsional force placed on sutures.
Copyright © 2012 Elsevier Ltd. All rights reserved.
Similar articles
-
Effect of saline conditions on the tensile strength of ophthalmic sutures.Ophthalmic Surg Lasers Imaging. 2011 Mar-Apr;42(2):148-51. doi: 10.3928/15428877-20101124-01. Epub 2010 Dec 1. Ophthalmic Surg Lasers Imaging. 2011. PMID: 21117584
-
Does plastic suture deformation induce gapping after tendon repair? A biomechanical comparison of different suture materials.J Biomech. 2016 Sep 6;49(13):2607-2612. doi: 10.1016/j.jbiomech.2016.05.023. Epub 2016 May 25. J Biomech. 2016. PMID: 27264620
-
Running sutures anchored with square knots are unreliable.Am J Surg. 2012 Sep;204(3):384-8. doi: 10.1016/j.amjsurg.2011.09.029. Epub 2012 Apr 10. Am J Surg. 2012. PMID: 22494881
-
A review of sutures and other skin closure materials.J Cosmet Laser Ther. 2010 Dec;12(6):296-302. doi: 10.3109/14764172.2010.538413. J Cosmet Laser Ther. 2010. PMID: 21142740 Review.
-
[Surgical basic skills: surgical sutures].Acta Med Port. 2011 Dec;24 Suppl 4:1051-6. Epub 2011 Dec 31. Acta Med Port. 2011. PMID: 22863517 Review. Portuguese.
Cited by
-
Surgical reconstruction of the urinary sphincter after traumatic longitudinal disruption.Case Rep Urol. 2014;2014:176073. doi: 10.1155/2014/176073. Epub 2014 Sep 2. Case Rep Urol. 2014. PMID: 25258694 Free PMC article.
-
[Surgical suture material-fundamentals].Oper Orthop Traumatol. 2023 Oct;35(5):298-316. doi: 10.1007/s00064-023-00812-y. Epub 2023 Aug 21. Oper Orthop Traumatol. 2023. PMID: 37603082 Free PMC article. German.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Miscellaneous