Microcomputer-based recording system for clinical electrophysiology
- PMID: 2276319
- DOI: 10.1007/BF00146555
Microcomputer-based recording system for clinical electrophysiology
Abstract
We developed a personal computer-based system for clinical electrophysiologic measurements. The computer interfaced with a commercially available A/D converter, a low-noise isolation preamplifier, filter circuits, pattern and Ganzfeld stimulators, and a hardcopy unit. Separate programs were developed for electroretinography (ERG), pattern ERG and simultaneous visual evoked potential (VEP), flash and pattern-shift VEP, and electro-oculographic measurements. The complete control of the applied hardware (eg, stimulus control, automatic gain, and filter selection) is a common feature of the computer programs. These programs provide oscilloscopic functions, overload protection, artifact elimination, averaging, automatic peak latency and amplitude determination, baseline correction, smoothing, and digital filtering. The results can be presented on matrix, laser printers, or digital plotters. The hardware components and the features of the driver software are demonstrated on normal and pathologic signals.
Similar articles
-
A personal computer-based visual evoked potential stimulus and recording system.Doc Ophthalmol. 1994;86(1):81-93. doi: 10.1007/BF01224630. Doc Ophthalmol. 1994. PMID: 7956688
-
[Digital data collection and analysis in clinical electrophysiology. A PC program for electrophysiology].Klin Monbl Augenheilkd. 1994 Jan;204(1):24-9. doi: 10.1055/s-2008-1035497. Klin Monbl Augenheilkd. 1994. PMID: 8152192 German.
-
Guidelines for calibration of stimulus and recording parameters used in clinical electrophysiology of vision. Calibration Standard Committee of the International Society for Clinical Electrophysiology of Vision (ISCEV).Doc Ophthalmol. 1998;95(1):1-14. doi: 10.1023/a:1001724411607. Doc Ophthalmol. 1998. PMID: 10189178
-
Current electrophysiology in ophthalmology: a review.Curr Opin Ophthalmol. 2012 Nov;23(6):497-505. doi: 10.1097/ICU.0b013e328359045e. Curr Opin Ophthalmol. 2012. PMID: 23047167 Review.
-
Clinical visual electrophysiology in China.Chin Med J (Engl). 1994 Jan;107(1):65-8. Chin Med J (Engl). 1994. PMID: 8187577 Review.
Cited by
-
Pattern electroretinogram and visual evoked potential amplitudes are influenced by different stimulus field sizes and scotomata.Doc Ophthalmol. 1993;83(2):139-49. doi: 10.1007/BF01206212. Doc Ophthalmol. 1993. PMID: 8334929
-
A personal computer-based visual evoked potential stimulus and recording system.Doc Ophthalmol. 1994;86(1):81-93. doi: 10.1007/BF01224630. Doc Ophthalmol. 1994. PMID: 7956688
-
Pattern electroretinogram, visual evoked potential and psychophysical functions in maculopathy.Doc Ophthalmol. 1995;90(3):229-45. doi: 10.1007/BF01203859. Doc Ophthalmol. 1995. PMID: 8846732
References
Publication types
MeSH terms
LinkOut - more resources
Medical