Effect of dental implant diameter on fatigue performance. Part I: mechanical behavior
- PMID: 22780911
- DOI: 10.1111/j.1708-8208.2012.00477.x
Effect of dental implant diameter on fatigue performance. Part I: mechanical behavior
Abstract
Aim: The purpose of this study was to evaluate the effect of the implants' diameter on the mechanical function and load-fatigue performance of dental implants.
Materials and methods: Three groups of implants with different diameters (3.3 mm, 3.75 mm and 5 mm), were tested under static and cyclic compressive loading. A total number of 15 implants for the static test and 112 implants for the cyclic-fatigue test. In the cyclic test, the machine ceased operating when the structure collapsed or when it reached 5 × 10(6) cycles without apparent failure. The load versus the number of cycles was plotted as curves for biomechanical analysis (S-N curve) for each implant diameter.
Results: The S-N curve plotted for the 5 mm implants showed classic fatigue behavior with a finite life region starting from 620N. The same was observed for the 3.75 mm diameter implants, with a finite life region starting below 620N. By contrast, the 3.3 mm diameter implants failed to show predictable fatigue behavior and a fatigue limit could not be defined.
Conclusions: The results of this study emphasize the importance of implant diameter on fatigue behavior. Narrow implants failed to show typical fatigue behavior which might be attributed to the implant design.
Keywords: S-N curve; cycles; load; normalized load; probability of fracture.
© 2012 Wiley Periodicals, Inc.
Similar articles
-
The effect of oral-like environment on dental implants' fatigue performance.Clin Oral Implants Res. 2014 Feb;25(2):e166-70. doi: 10.1111/clr.12084. Epub 2012 Dec 20. Clin Oral Implants Res. 2014. PMID: 23278444
-
Load limit of mini-implants with reduced abutment height based on fatigue fracture resistance: experimental and finite element study.Int J Oral Maxillofac Implants. 2015 Jan-Feb;30(1):e10-6. doi: 10.11607/jomi.3653. Epub 2014 Dec 5. Int J Oral Maxillofac Implants. 2015. PMID: 25506647
-
Influence of the height of the external hexagon and surface treatment on fatigue life of commercially pure titanium dental implants.Int J Oral Maxillofac Implants. 2009 Jul-Aug;24(4):583-90. Int J Oral Maxillofac Implants. 2009. PMID: 19885397
-
Impact of dental implant diameter on the efficiency of fatigue: A systematic review analysis.J Pak Med Assoc. 2021 Jun;71(6):1648-1654. doi: 10.47391/JPMA.1040. J Pak Med Assoc. 2021. PMID: 34111090
-
Optimized Planning and Evaluation of Dental Implant Fatigue Testing: A Specific Software Application.Biology (Basel). 2020 Oct 31;9(11):372. doi: 10.3390/biology9110372. Biology (Basel). 2020. PMID: 33142807 Free PMC article. Review.
Cited by
-
The Effect of Implantoplasty on the Fatigue Behavior and Corrosion Resistance in Titanium Dental Implants.Materials (Basel). 2024 Jun 15;17(12):2944. doi: 10.3390/ma17122944. Materials (Basel). 2024. PMID: 38930312 Free PMC article.
-
Fracture resistance after implantoplasty in three implant-abutment connection designs.Med Oral Patol Oral Cir Bucal. 2020 Sep 1;25(5):e691-e699. doi: 10.4317/medoral.23700. Med Oral Patol Oral Cir Bucal. 2020. PMID: 32683385 Free PMC article.
-
Influence of Narrow Titanium Dental Implant Diameter on Fatigue Behavior: A Comparison between Unitary and Splinted Implants.J Clin Med. 2024 Mar 13;13(6):1632. doi: 10.3390/jcm13061632. J Clin Med. 2024. PMID: 38541858 Free PMC article.
-
Implantoplasty Improves Clinical Parameters over a 2-Year Follow-Up: A Case Series.Medicina (Kaunas). 2022 Jan 12;58(1):113. doi: 10.3390/medicina58010113. Medicina (Kaunas). 2022. PMID: 35056421 Free PMC article.
-
Current trends in dental implants.J Korean Assoc Oral Maxillofac Surg. 2014 Apr;40(2):50-60. doi: 10.5125/jkaoms.2014.40.2.50. Epub 2014 Apr 28. J Korean Assoc Oral Maxillofac Surg. 2014. PMID: 24868501 Free PMC article. Review.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources