Comparative evaluation of roughness of titanium surfaces treated by different hygiene instruments
- PMID: 22803010
- PMCID: PMC3395000
- DOI: 10.5051/jpis.2012.42.3.88
Comparative evaluation of roughness of titanium surfaces treated by different hygiene instruments
Abstract
Purpose: The use of appropriate instruments to clean surfaces with minimal change, is critical for the successful maintenance of a dental implant. However, there is no consensus about the type and methodology for such instruments. The aim of this study was to characterize changes in the roughness of titanium surfaces treated by various scaling instruments.
Methods: Thirty-seven identical disks (5 mm in diameter) were investigated in this study. The specimens were divided into eight groups according to the types of instrumentation and the angle of application. Ultrasonic scaling systems were applied on a titanium disk to simulate standard clinical conditions. The equipment included a piezoelectric ultrasonic scaler with a newly developed metallic tip (NS group), a piezoelectric ultrasonic scaler with a conventional tip (CS group), a piezoelectric root planer ultrasonic scaler with a conventional tip (PR group), and a plastic hand curette (PH group). In addition, the sites treated using piezoelectric ultrasonic scaler systems were divided two sub-groups: 15 and 45 degrees. The treated titanium surfaces were observed by scanning electron microscopy (SEM), and the average surface roughness (Ra) and mean roughness profile depth (Rz) were measured with a profilometer.
Results: SEM no significant changes in the titanium surfaces in the NS group, regardless of the angle of application. The PH group also showed no marked changes to the titanium surface, although some smoothening was observed. All CS and PR sites lost their original texture and showed irregular surfaces in SEM analysis. The profilometer analysis demonstrated that the roughness values (Ra and Rz) of the titanium surfaces increased in all, except the PH and NS groups, which showed roughness decreases relative to the untreated control group. The Ra value differed significantly between the NS and PR groups (P<0.05).
Conclusions: The results of this study indicated that changes in or damage to titanium surfaces might be more affected by the hardness of the scaler tip than by the application method. Within the limitations of this study, the newly developed metallic scaler tip might be especially suitable for peri-implant surface decontamination, due to its limited effects on the titanium surface.
Keywords: Dental implants; Dental instruments; Peri-implantitis; Periodontal debridement.
Conflict of interest statement
No potential conflict of interest relevant to this article was reported.
Figures





Similar articles
-
Effects on the Titanium Implant Surface by Different Hygiene Instrumentations: A Narrative Review.Cureus. 2022 Oct 30;14(10):e30884. doi: 10.7759/cureus.30884. eCollection 2022 Oct. Cureus. 2022. PMID: 36465763 Free PMC article. Review.
-
The treatment of uncoated and titanium nitride-coated abutments with different instruments.Int J Oral Maxillofac Implants. 2004 Mar-Apr;19(2):232-8. Int J Oral Maxillofac Implants. 2004. PMID: 15101595
-
Effects of ultrasonic scaler tips and toothbrush on titanium disc surfaces evaluated with confocal microscopy.J Craniofac Surg. 2012 Sep;23(5):1552-8. doi: 10.1097/SCS.0b013e31825e3ba6. J Craniofac Surg. 2012. PMID: 22976659
-
An in vitro evaluation of zirconia surface roughness caused by different scaling methods.J Prosthet Dent. 2010 May;103(5):283-7. doi: 10.1016/S0022-3913(10)60059-5. J Prosthet Dent. 2010. PMID: 20416411
-
Implant surface alterations from a nonmetallic ultrasonic tip.J West Soc Periodontol Periodontal Abstr. 1998;46(3):69-73. J West Soc Periodontol Periodontal Abstr. 1998. PMID: 10597152 Review.
Cited by
-
Definition, etiology, prevention and treatment of peri-implantitis--a review.Head Face Med. 2014 Sep 3;10:34. doi: 10.1186/1746-160X-10-34. Head Face Med. 2014. PMID: 25185675 Free PMC article. Review.
-
Bacterial species associated with peri-implant disease - a literature review.Germs. 2023 Dec 31;13(4):352-361. doi: 10.18683/germs.2023.1405. eCollection 2023 Dec. Germs. 2023. PMID: 38361546 Free PMC article. Review.
-
Effects on the Titanium Implant Surface by Different Hygiene Instrumentations: A Narrative Review.Cureus. 2022 Oct 30;14(10):e30884. doi: 10.7759/cureus.30884. eCollection 2022 Oct. Cureus. 2022. PMID: 36465763 Free PMC article. Review.
-
An In Vivo Investigation of Non-Metallic vs. Metallic Hand Scalers on Zirconia Implant-Supported Crowns: A Year-Long Analysis of Peri-Implant Maintenance.J Funct Biomater. 2023 Dec 23;15(1):9. doi: 10.3390/jfb15010009. J Funct Biomater. 2023. PMID: 38248676 Free PMC article.
-
Decontamination methods to restore the biocompatibility of contaminated titanium surfaces.J Periodontal Implant Sci. 2019 Jun 25;49(3):193-204. doi: 10.5051/jpis.2019.49.3.193. eCollection 2019 Jun. J Periodontal Implant Sci. 2019. PMID: 31285943 Free PMC article.
References
-
- Lindhe J, Meyle J Group D of European Workshop on Periodontology. Peri-implant diseases: Consensus Report of the Sixth European Workshop on Periodontology. J Clin Periodontol. 2008;35(8 Suppl):282–285. - PubMed
-
- Mombelli A, van Oosten MA, Schurch E, Jr, Land NP. The microbiota associated with successful or failing osseointegrated titanium implants. Oral Microbiol Immunol. 1987;2:145–151. - PubMed
-
- Leonhardt A, Renvert S, Dahlen G. Microbial findings at failing implants. Clin Oral Implants Res. 1999;10:339–345. - PubMed
-
- Costerton JW, Stewart PS, Greenberg EP. Bacterial biofilms: a common cause of persistent infections. Science. 1999;284:1318–1322. - PubMed
-
- Lamont RJ, Jenkinson HF. Subgingival colonization by Porphyromonas gingivalis. Oral Microbiol Immunol. 2000;15:341–349. - PubMed
LinkOut - more resources
Full Text Sources
Medical
Research Materials
Miscellaneous