Renal sympathetic denervation using an irrigated radiofrequency ablation catheter for the management of drug-resistant hypertension
- PMID: 22814781
- DOI: 10.1016/j.jcin.2012.01.027
Renal sympathetic denervation using an irrigated radiofrequency ablation catheter for the management of drug-resistant hypertension
Abstract
Objectives: This study sought to assess whether renal sympathetic denervation (RSDN) can be achieved using an off-the-shelf saline-irrigated radiofrequency ablation (RFA) catheter typically employed for cardiac tissue ablation.
Background: RSDN using a specialized solid-tip RFA catheter has recently been demonstrated to safely reduce systemic blood pressure in patients with refractory hypertension. For cardiac tissue ablation, RFA technology has evolved from nonirrigated to saline-irrigated ablation electrodes to improve both safety and effectiveness.
Methods: Ten patients with resistant hypertension underwent renal angiography, followed by bilateral RSDN with a saline-irrigated RFA catheter. Ambulatory blood pressure recordings (24 h) were obtained at baseline, 1, 3, and 6 months after the procedure. Repeat renal angiography was performed during follow-up to assess for arterial stenosis or aneurysm. In 5 patients, pre- and post-procedural serum measures of renal function and sympathetic activity were obtained: aldosterone; metanephrine; normetanephrine; plasma renin activity; and creatinine.
Results: Over a 6-month period: 1) the systolic/diastolic blood pressure decreased by -21/-11 mm Hg; 2) all patients experienced a decrease in systolic blood pressure of at least 10 mm Hg (range 10 to 40 mm Hg); 3) there was no evidence of renal artery stenosis or aneurysm at repeat angiography; and 4) there was a significant decrease in metanephrine (-12 ± 4, p = 0.003), normetanephrine (-18 ± 4, p = 0.0008), and aldosterone levels (-60 ± 33 ng/l, p = 0.02) at 3 months. There was no significant change in plasma renin activity (-0.2 mg/l/hod, p = 0.4). There was no significant change in serum creatinine (-1 mmol/l, p = 0.4).
Conclusions: These data provide the proof-of-principle that RSDN can be performed using an off-the-shelf saline-irrigated RFA catheter.
Copyright © 2012 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.
Comment in
-
Renal denervation using an irrigated radiofrequency ablation catheter for management of drug-resistant hypertension: a demonstrated value?JACC Cardiovasc Interv. 2012 Jul;5(7):766-8. doi: 10.1016/j.jcin.2012.02.019. JACC Cardiovasc Interv. 2012. PMID: 22814782 No abstract available.
-
Cardiac ablation and renal denervation systems have distinct purposes and different technical requirements.JACC Cardiovasc Interv. 2013 Mar;6(3):314. doi: 10.1016/j.jcin.2012.12.005. JACC Cardiovasc Interv. 2013. PMID: 23517846 No abstract available.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Research Materials
Miscellaneous