Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2013 Feb;15(1):37-48.
doi: 10.1007/s10544-012-9685-0.

A MEMS electrochemical bellows actuator for fluid metering applications

Affiliations

A MEMS electrochemical bellows actuator for fluid metering applications

Roya Sheybani et al. Biomed Microdevices. 2013 Feb.

Abstract

We present a high efficiency wireless MEMS electrochemical bellows actuator capable of rapid and repeatable delivery of boluses for fluid metering and drug delivery applications. Nafion®-coated Pt electrodes were combined with Parylene bellows filled with DI water to form the electrolysis-based actuator. The performance of actuators with several bellows configurations was compared for a range of applied currents (1-10 mA). Up to 75 boluses were delivered with an average pumping flow rate of 114.40 ± 1.63 μL/min. Recombination of gases into water, an important factor in repeatable and reliable actuation, was studied for uncoated and Nafion®-coated actuators. Real-time pressure measurements were conducted and the effects of temperature, physiological back pressure, and drug viscosity on delivery performance were investigated. Lastly, we present wireless powering of the actuator using a class D inductive powering system that allowed for repeatable delivery with less than 2 % variation in flow rate values.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Illustration of operation concept of electrochemical bellows actuator. Actuator is shown housed within a reservoir to illustrate the case of fluid metering through an attached catheter.
Figure 2
Figure 2
Illustration detailing the electrode and bellows fabrication processes and bellows actuator assembly. A 2 convolution bellows is shown.
Figure 3
Figure 3
Photograph of: (a) Pt electrolysis electrode, (b) 2 convolution Parylene bellows, (c) assembled bellows actuator.
Figure 4
Figure 4
Circuit schematic for class D wireless inductive powering system.
Figure 5
Figure 5
(a) Real-time pressure measurements of a 2 convolution Nafion®-coated bellows actuator under constant current application; (b) Slope values for real-time pressure vs. time for different current values.
Figure 6
Figure 6
Comparison between recombined volume for uncoated and coated electrodes: (a) current controlled (10 mA) delivery for different ON/OFF times (Sheybani, Gensler et al. 5-9 June 2011) and (b) time controlled (2 min) flow rate delivery at different applied currents.
Figure 7
Figure 7
Comparison between rapid-fire bolus delivery using uncoated and coated 2 convolution bellows actuators (3 boluses at 10 mA, 15 sec ON/10 sec OFF, separated by 5 min OFF cycles (Sheybani, Gensler et al. 2011)).
Figure 8
Figure 8
Bolus delivery with a Nafion® coated, 2 convolution bellows actuator (at 10 mA current, 15 sec/1 min ON/OFF, modified from (Sheybani, Gensler et al. 2011)).
Figure 9
Figure 9
Bolus delivery with a Nafion® coated, 2 convolution bellows actuator (at 10mA current, 2 sec/1 min ON/OFF), inset: close-up for first 10 boluses delivered.
Figure 10
Figure 10
Bolus delivery using an actuator with an in-line check valve (at 10 mA current, 15 sec ON/ 60 sec OFF). After 15 boluses, the actuator is turned off. Inset shows average accumulated volume for 15 boluses.
Figure 11
Figure 11
Flow delivery results for of different viscosities of propylene glycol (at 8 mA constant current).
Figure 12
Figure 12
Current controlled flow delivery of cocaine (in 0.9 N saline) loaded in a pump at different concentrations (at 8 mA, 1.5 convolution bellows, modified from (Sheybani, Gensler et al. 2011)).
Figure 13
Figure 13
Flow delivery results for a range of physiologically relevant back pressures (at 5mA constant current).

Similar articles

Cited by

References

    1. Amirouche F, Yu Z, et al. Current micropump technologies and their biomedical applications. Microsystem Technologies. 2009;15(5):647–666.
    1. Bohm S, Timmer B, et al. Closed-loop controlled electrochemically actuated micro-dosing system. Journal of Micromechanics and Microengineering. 2000;10:498–504. Compendex.
    1. Broka K, Ekdunge P. Oxygen and hydrogen permeation properties and water uptake of Nafion 117 membrane and recast film for PEM fuel cell. Journal of Applied Electrochemistry. 1997;27:117–123. Copyright 1997, IEE.
    1. Bruguerolle B, Labrecque G. Rhythmic pattern in pain and their chronotherapy. Advanced Drug Delivery Reviews. 2007;59(9-10):883–895. - PubMed
    1. Cameron CG, Freund MS. Electrolytic actuators: Alternative, high-performance, material-based devices. Proceedings of the National Academy of Sciences of the United States of America. 2002;99(12):7827–7831. - PMC - PubMed

Publication types

Substances

LinkOut - more resources