Radiation and osteosynthesis. Dosimetry on an irradiation phantom
- PMID: 2283402
- DOI: 10.1016/s1010-5182(05)80057-5
Radiation and osteosynthesis. Dosimetry on an irradiation phantom
Abstract
An irradiation phantom was used to measure dose increases using the backscatter of different materials (titanium, steel, lead, aluminium). Telecobalt-60 and 8-MV photons were used for the irradiation. The irradiation dose was measured by means of a defined X-ray film blackening. The most important parameter was to find out whether under simulated conditions, undesirable hot spots occur. We were able to demonstrate that a 12.5 to 16% increase in the radiation dose can be observed for titanium and steel at a distance of 0.45 mm from the metal specimen. A comparison between titanium and steel did not demonstrate a relevant advantage for titanium. Therefore, adjuvant percutaneous radiation therapy should not have any influence on the life of the implant, if the soft tissue layer is of sufficient thickness. The available literature is reviewed and presented in tables.
Similar articles
-
[Tumor irradiation after plate osteosynthesis. Dose determination on radiological phantom].Dtsch Z Mund Kiefer Gesichtschir. 1989 May-Jun;13(3):165-71. Dtsch Z Mund Kiefer Gesichtschir. 1989. PMID: 2625006 German.
-
Backscatter radiation at tissue-titanium interfaces. Analyses of biological effects from 60Co and protons.Acta Oncol. 1991;30(7):859-66. doi: 10.3109/02841869109091835. Acta Oncol. 1991. PMID: 1662525
-
[Thermoluminescent dosimetry studies on absorptive behavior of bones without and with metal implants].Strahlenther Onkol. 1986 Dec;162(12):779-84. Strahlenther Onkol. 1986. PMID: 3101213 German.
-
[Susceptibility to infections and behavior of stainless steel : Comparison with titanium implants in traumatology].Unfallchirurg. 2017 Feb;120(2):110-115. doi: 10.1007/s00113-016-0300-3. Unfallchirurg. 2017. PMID: 28070629 Review. German.
-
An improved head-and-neck phantom for radiation dosimetry.Oral Surg Oral Med Oral Pathol. 1989 Mar;67(3):338-46. doi: 10.1016/0030-4220(89)90367-8. Oral Surg Oral Med Oral Pathol. 1989. PMID: 2648244 Review.
Cited by
-
Postoperative irradiation after implant placement: A pilot study for prosthetic reconstruction.J Adv Prosthodont. 2016 Oct;8(5):363-371. doi: 10.4047/jap.2016.8.5.363. Epub 2016 Oct 21. J Adv Prosthodont. 2016. PMID: 27826386 Free PMC article.
-
The Carbofix™ "Piccolo Proximal femur nail": A new perspective for treating proximal femur lesion. A technique report.J Orthop. 2016 Jul 6;13(4):343-6. doi: 10.1016/j.jor.2016.07.001. eCollection 2016 Dec. J Orthop. 2016. PMID: 27436924 Free PMC article.
-
Effects of titanium mesh implant on dosimetry during Gamma Knife radiosurgery.J Appl Clin Med Phys. 2012 Sep 6;13(5):3833. doi: 10.1120/jacmp.v13i5.3833. J Appl Clin Med Phys. 2012. PMID: 22955648 Free PMC article.